01.11.2012 Views

Optimizing the Analysis of Volatile Organic Compounds

Optimizing the Analysis of Volatile Organic Compounds

Optimizing the Analysis of Volatile Organic Compounds

SHOW MORE
SHOW LESS

Create successful ePaper yourself

Turn your PDF publications into a flip-book with our unique Google optimized e-Paper software.

Connect <strong>the</strong> capillary column, 0.25mm ID or smaller, to <strong>the</strong> injection port, but not to <strong>the</strong><br />

MS source, and condition <strong>the</strong> column. When <strong>the</strong> column has been conditioned, remove<br />

50cm from <strong>the</strong> detector end <strong>of</strong> <strong>the</strong> column to ensure complete removal <strong>of</strong> siloxanes and<br />

o<strong>the</strong>r potential contaminants. Then, with <strong>the</strong> MS still turned <strong>of</strong>f, insert <strong>the</strong> column end into<br />

<strong>the</strong> MS source. Cut <strong>the</strong> column several centimeters from <strong>the</strong> connection to <strong>the</strong> injection port.<br />

Use septa to cap <strong>the</strong> short length <strong>of</strong> column that is left in <strong>the</strong> injection port and <strong>the</strong> new,<br />

unconnected inlet end <strong>of</strong> <strong>the</strong> column. Also cap <strong>the</strong> split vent and septum purge vent lines on<br />

<strong>the</strong> GC (Figure 38). Perform a pressure decay test on <strong>the</strong> injection port by setting <strong>the</strong> pressure<br />

to 30psi, <strong>the</strong>n shutting <strong>of</strong>f <strong>the</strong> gas supply. The pressure should remain constant for at<br />

least 1 minute. If <strong>the</strong> pressure drops in less than 1 minute, turn on <strong>the</strong> gas supply and begin<br />

leak checking, using an electronic leak detector, such as <strong>the</strong> Restek Leak Detective TM II (cat.<br />

#20413, page 36).<br />

Once you have confirmed <strong>the</strong> GC system is leak-tight, and while <strong>the</strong> injection port end <strong>of</strong><br />

<strong>the</strong> column is capped and <strong>the</strong>re is no flow in <strong>the</strong> column, evacuate <strong>the</strong> MS and record <strong>the</strong><br />

source pressure in your maintenance logbook. After several hours equilibration, perform <strong>the</strong><br />

instrument leak-check using PFTBA (see MS Calibration and Tuning, page 32). If a leak is<br />

present, draw 500µL <strong>of</strong> methanol into a syringe and apply drops <strong>of</strong> this solvent on areas<br />

where leaks might be suspected, while scanning for mass 31. Alternatively, ba<strong>the</strong> <strong>the</strong>se areas<br />

with argon gas and scan for m/z 40. The brass source nut is <strong>the</strong> primary place for leaks in an<br />

MS, this nut should be replaced every time <strong>the</strong> column is changed (cat. #20643, page 35).<br />

O<strong>the</strong>r areas to examine include <strong>the</strong> rubber seals and <strong>the</strong> PFTBA vial.<br />

After you confirm <strong>the</strong> MS is leak-free, quickly install <strong>the</strong> inlet end <strong>of</strong> <strong>the</strong> column into <strong>the</strong><br />

injection port. The MS will draw air into <strong>the</strong> source during this connection time; after ano<strong>the</strong>r<br />

2-hour equilibration <strong>the</strong> MS is ready for tuning/leak checking using PFTBA.<br />

Figure 37.<br />

Cl and Br are examples <strong>of</strong> A+2 elements (isotopes vary by 2amu or more).<br />

Cl Cl2 Br<br />

Br2<br />

Element Mass % Abundance Mass % Abundance<br />

carbon 12 100 13 1.1<br />

nitrogen 14 100 15 0.37<br />

chlorine 35 100 37 32.5<br />

bromine 79 100 81 98<br />

oxygen 16 100 18 0.2<br />

Table V.<br />

Fragment ions can <strong>of</strong>fer clues to compound composition.<br />

Compound Class Fragment Ions<br />

Aldehydes, amides, amines 44, 58, 72, 59, 30<br />

Aliphatic hydrocarbons 43, 57, 71, 85, 99<br />

Alkylbenzenes 104, 91<br />

Aromatic hydrocarbons 39, 50, 51, 52, 63, 65, 76, 77, 91<br />

Fluorine-containing 50, 69<br />

Methacrylates 41, 69<br />

Methyl ketones 43, 58<br />

Oxygen-containing 31,45,59,73<br />

Sulfur-containing 47,61<br />

Unsaturated hydrocarbons 41, 55, 69<br />

35<br />

Figure 38.<br />

To check for leaks, cap <strong>the</strong> split vent and<br />

septum purge vent lines on <strong>the</strong> GC and <strong>the</strong><br />

column, <strong>the</strong>n shut <strong>of</strong>f <strong>the</strong> gas supply.<br />

System pressure should remain constant.<br />

Cap both<br />

MSD Source Nut<br />

(Similar to Agilent part # 05988-20066.)<br />

Description qty. cat.#<br />

(Detector) MSD Source Nut 2-pk. 20643<br />

www.restekcorp.com

Hooray! Your file is uploaded and ready to be published.

Saved successfully!

Ooh no, something went wrong!