01.11.2012 Views

Optimizing the Analysis of Volatile Organic Compounds

Optimizing the Analysis of Volatile Organic Compounds

Optimizing the Analysis of Volatile Organic Compounds

SHOW MORE
SHOW LESS

Create successful ePaper yourself

Turn your PDF publications into a flip-book with our unique Google optimized e-Paper software.

32<br />

Figure 31.<br />

An open split interface or a jet separator<br />

will reduce <strong>the</strong> gas flow entering a mass<br />

spectrometer vacuum system.<br />

carrier<br />

gas<br />

make-up<br />

gas<br />

purge in<br />

open split interface<br />

jet separator<br />

vacuum<br />

www.restekcorp.com<br />

to MSD<br />

1:10<br />

split<br />

purge out<br />

to<br />

MSD<br />

MS Calibration and Tuning<br />

Calibration allows <strong>the</strong> correct identification <strong>of</strong> masses, whereas tuning adjusts <strong>the</strong> intensity<br />

and peakwidths for masses. The MS is calibrated by adjusting <strong>the</strong> DC/RF frequency so that<br />

mass axis points are aligned with expected mass fragments <strong>of</strong> known spectra. Tuning<br />

ensures that target compounds analyzed on <strong>the</strong> MS will have <strong>the</strong> same distribution (pattern)<br />

<strong>of</strong> ions, and peak widths for ions will be narrow enough that adjacent mass peaks will not<br />

overlap. A compound widely used for calibrating and tuning MS systems is perfluorotributylamine<br />

(PFTBA or FC43). Modern instruments introduce PFTBA into <strong>the</strong> ion source during<br />

<strong>the</strong> autotune procedure. The instrument s<strong>of</strong>tware adjusts <strong>the</strong> MS parameters to match <strong>the</strong><br />

known fragmentation pattern for PFTBA. The ion <strong>of</strong> greatest abundance in <strong>the</strong> spectrum is<br />

mass 69; <strong>the</strong> relative abundances <strong>of</strong> masses 131 and 219 are roughly 50% <strong>of</strong> that for mass<br />

69 (Figure 32). In analyses <strong>of</strong> volatiles, mass 502 is less important because its relative abundance<br />

is 1% <strong>of</strong> <strong>the</strong> mass 69 value. Low peak heights or a loss <strong>of</strong> mass 502 generally indicate<br />

a cleanliness problem at <strong>the</strong> source.<br />

Specific Tuning Requirements: 4-Brom<strong>of</strong>luorobenzene: After <strong>the</strong> system is calibrated and<br />

tuned, using PFTBA, a 50ng solution <strong>of</strong> 4-brom<strong>of</strong>luorobenzene (BFB) is introduced. BFB<br />

usually is introduced by injection through <strong>the</strong> GC injection port but, alternatively, it can be<br />

purged from a water blank. Abundance criteria for BFB are listed in Table IV.<br />

If tuning with BFB fails under criteria acceptable for PFTBA, decrease <strong>the</strong> relative abundance<br />

<strong>of</strong> masses 131 and 219 to 30% <strong>of</strong> mass 69 by adjusting <strong>the</strong> entrance lens. If necessary,<br />

slightly decrease <strong>the</strong> repeller voltage. This procedure targets <strong>the</strong> ions from mass 173 through<br />

mass 177. A second tuning failure with BFB may dictate recalibration and tuning with<br />

PFTBA. Ion ratios for BFB should be checked every 12 hours. As long as results meet <strong>the</strong><br />

specifications in Table IV, no fur<strong>the</strong>r calibration or tuning is required.<br />

Poor tuning can significantly affect <strong>the</strong> sensitivity <strong>of</strong> <strong>the</strong> MS. Figure 33 shows spectra for a<br />

sample analyzed twice, first after a failing PFTBA tuning with mass 131 as <strong>the</strong> base peak<br />

(Figure 33, A), <strong>the</strong>n after a passing tuning (Figure 33, B). The second analysis exhibits a<br />

three-fold increase in sensitivity.<br />

Figure 32.<br />

Perfluorotributylamine (PFTBA) fragmentation, using a quadrupole mass spectrometer.<br />

69<br />

100<br />

80<br />

60<br />

40<br />

20<br />

0<br />

131<br />

219<br />

50 100 150 200 250 300 350 400 450 500 550 600<br />

Table IV.<br />

US EPA ion abundance criteria for 4-brom<strong>of</strong>luorobenzene (BFB).<br />

Mass/Charge Ratio Relative Abundance Criterion<br />

50 15-50% <strong>of</strong> mass 95<br />

75 30-80% <strong>of</strong> mass 95<br />

95 Base peak, 100% relative abundance<br />

96 5-9% <strong>of</strong> mass 95<br />

173 50% <strong>of</strong> mass 95<br />

175 5-9% <strong>of</strong> mass 174<br />

176 >95% but

Hooray! Your file is uploaded and ready to be published.

Saved successfully!

Ooh no, something went wrong!