01.11.2012 Views

Optimizing the Analysis of Volatile Organic Compounds

Optimizing the Analysis of Volatile Organic Compounds

Optimizing the Analysis of Volatile Organic Compounds

SHOW MORE
SHOW LESS

Create successful ePaper yourself

Turn your PDF publications into a flip-book with our unique Google optimized e-Paper software.

Detection Systems<br />

VOCs can be analyzed using a variety <strong>of</strong> detection systems, including GC/MS, GC/PID, and<br />

GC/ELCD. Here, we discuss consequences <strong>of</strong> using each <strong>of</strong> <strong>the</strong>se systems, and present tips<br />

for maintenance and troubleshooting.<br />

Column Configurations<br />

Single Column: Environmental engineers characterize a contaminated site using MS or<br />

dual-column GC, or <strong>the</strong>y might monitor <strong>the</strong> site solely with single-column GC methods.<br />

Injections <strong>of</strong> standards on a single column, delivered to <strong>the</strong> detector, can provide tentative<br />

identification and quantification. Retention times for analytes listed in a given method are<br />

established by injecting a check standard containing all <strong>of</strong> <strong>the</strong> target compounds. Retention<br />

times for analytes in site samples are compared against retention times for <strong>the</strong> standard, to<br />

verify if unknown compounds match known targets. A single column configuration works<br />

well with characterized samples, but retention times are not unique for every analyte, especially<br />

for early-eluting compounds that spend little time in <strong>the</strong> stationary phase (e.g., Freon ®<br />

compounds). In environmental laboratories coelutions from non-target compounds also are<br />

very common, creating very complex chromatograms that are difficult to interpret using a<br />

single-column design.<br />

Dual Columns: In a dual-column configuration, <strong>the</strong> sample passes through a fused silica<br />

guard column, <strong>the</strong>n is split between two analytical columns <strong>of</strong> differing selectivity.<br />

Standards are injected to establish retention times on both columns simultaneously. One disadvantage<br />

to this configuration is <strong>the</strong> 50% loss in sensitivity resulting from splitting <strong>the</strong><br />

sample. This loss can be overcome by increasing <strong>the</strong> sample volume or by optimizing <strong>the</strong><br />

detector. Flow rates for <strong>the</strong> two columns should agree within 20% because uneven splitting<br />

will fur<strong>the</strong>r affect sensitivity.<br />

Detector Configurations<br />

Detectors can be connected in parallel, in series, or in tandem, to double <strong>the</strong> amount <strong>of</strong><br />

information about <strong>the</strong> sample.<br />

Parallel System: In a parallel system <strong>the</strong> sample is split equally between <strong>the</strong> two detectors,<br />

allowing both detectors to be destructive (e.g., ELCD/FID). This detection system works<br />

well but is unsuitable for a dual-column analytical configuration because <strong>the</strong> sample already<br />

will have been split between <strong>the</strong> two columns.<br />

Series System: Series detection involves connecting two detectors in sequence, using a short<br />

length <strong>of</strong> deactivated metal or fused silica tubing. The sample passes through <strong>the</strong> first detector,<br />

which must be non-destructive (e.g., PID), <strong>the</strong>n through <strong>the</strong> second detector. This produces<br />

two sets <strong>of</strong> information about <strong>the</strong> sample with no loss in sensitivity because <strong>the</strong> sample<br />

volume does not change from <strong>the</strong> first detector to <strong>the</strong> second. The only disadvantage is<br />

dead volume, which can broaden <strong>the</strong> peaks. Minimize dead volume by minimizing <strong>the</strong><br />

length and ID <strong>of</strong> <strong>the</strong> line connecting <strong>the</strong> detectors.<br />

Tandem System: The tandem configuration connects two detectors without <strong>the</strong> dead volume<br />

associated with a series system. The non-destructive first detector is <strong>the</strong> base for <strong>the</strong> second<br />

detector. The units can be connected to a single detector port on one GC. This makes it possible<br />

to use a dual-column configuration, with each column connecting to tandem detectors,<br />

producing four sets <strong>of</strong> data per analysis. This approach is used in EPA Method 8021. 9<br />

Detectors: Method requirements determine <strong>the</strong> choice <strong>of</strong> detector(s). The current shift<br />

toward analysis by performance-based criteria makes it possible to use detection o<strong>the</strong>r than<br />

that listed in a method if it can be shown that performance is similar to, or better than, what<br />

would be attained by following <strong>the</strong> guidelines in <strong>the</strong> method. The most common GC detectors<br />

are <strong>the</strong> PID, <strong>the</strong> FID, and <strong>the</strong> ELCD. GC/MS eliminates <strong>the</strong> need for a confirmation<br />

column.<br />

9. R.D. Braun. Introduction to Instrumental <strong>Analysis</strong>. McGraw-Hill Book Company. New York. 1987.<br />

pp. 915-916.<br />

23<br />

www.restekcorp.com

Hooray! Your file is uploaded and ready to be published.

Saved successfully!

Ooh no, something went wrong!