Figure 14.<br />
Configuring your GC is simple! Prepare <strong>the</strong> lines by cutting <strong>the</strong>m as shown. (Agilent 6880)<br />
Attaches to Electronic Pressure<br />
Control (EPC) Unit <strong>of</strong><br />
Chromatograph<br />
Septum Purge<br />
Split Line<br />
To P&T<br />
Carrier Gas<br />
Cut Lines, Add Brass Nuts<br />
Figure 15.<br />
Carrier gas flow is adjusted through <strong>the</strong> injection port regulator. Carrier gas sweeps<br />
analytes from <strong>the</strong> trap, through <strong>the</strong> transfer line, onto <strong>the</strong> column. (Tekmar 3100)<br />
Heated Transfer<br />
Line (To GC<br />
Injection Port)<br />
Gas Lines<br />
Figure 16.<br />
Reconnect <strong>the</strong> weldment lines to <strong>the</strong> GC and check for leaks.<br />
Be sure to set up <strong>the</strong> lines correctly. (Agilent 6890)<br />
From GC<br />
Carrier Gas<br />
Split Vent<br />
Septum Purge<br />
packed columns, we recommend using a Vu-Tight ® inlet liner (cat. #20342, page 18). The<br />
wide-bore capillary column is sealed into <strong>the</strong> tapered restriction in <strong>the</strong> liner, ensuring direct<br />
transfer <strong>of</strong> <strong>the</strong> sample to <strong>the</strong> column. In addition, <strong>the</strong> Vu-Tight ® inlet liner allows visual<br />
inspection <strong>of</strong> <strong>the</strong> column-to-liner seal.<br />
Low-Volume Injectors: A low-volume injector (LVI) will reduce dead volume, yet allow<br />
limited manual injections. Such a system can be used to convert a packed column or a capillary<br />
split/splitless injection port for purge and trap applications. The internal volume <strong>of</strong> <strong>the</strong><br />
injector is significantly reduced, compared to a conventional injection port, which improves<br />
sample transfer from <strong>the</strong> purge and trap system. The purge and trap transfer line is connected<br />
to <strong>the</strong> LVI, and <strong>the</strong> wide-bore column is connected at <strong>the</strong> base <strong>of</strong> <strong>the</strong> injector. A septum in <strong>the</strong><br />
LVI allows manual injections if needed.<br />
Low Volume Injectors<br />
19<br />
Description cat.#<br />
Low-Volume Injector for Agilent<br />
Split/Splitless GC Inlets 21692<br />
Low-Volume Injector for Agilent<br />
5890 Septum Packed Purge Port 21698<br />
Low-Volume Injector for Varian<br />
Split/Splitless GC Inlets 21693<br />
For descriptions <strong>of</strong> low-volume injectors,<br />
see page 63.<br />