Placement of the ceramic insert<br />
The procedure for the trial insert is the same<br />
as described under “Optionally: trial reduction<br />
using trial insert”.<br />
Before insertion it must be ensured that the<br />
inner surface of the metal shell is dry and<br />
free of any debris.<br />
Remove the BIOLOX ® forte ceramic insert<br />
with the premounted CeraLock ® insertion aid<br />
from its packaging. Remove the transparent<br />
Plexiglas triggering protection and dispose<br />
of it.<br />
Place the CeraLock ® insertion aid on the front<br />
face of the metal shell so that all 3 arms lie<br />
on the edge and the BIOLOX ® forte ceramic<br />
insert cannot tilt.<br />
Exert force on the CeraLock ® plateau and<br />
press the BIOLOX ® forte ceramic insert home<br />
into the shell.<br />
The instrument will have now released the<br />
BIOLOX ® forte insert. Remove the CeraLock ®<br />
insertion aid and dispose of it. It is intended<br />
for single usage only and must not be resterilized.<br />
The correct position of the insert in the metal<br />
shell must be checked. Using the tip of the<br />
fingers, touch the 2 front faces of the insert<br />
and metal shell. If the position of the insert<br />
over its entire circumference is level to the<br />
metal edge, then it is correctly positioned in<br />
the shell.<br />
14