Optionally: trial reduction using trial<br />

insert (standard 130826–130839/hooded<br />

130850–130857)<br />

Inner section<br />

The trial stem is inserted in the femur after<br />

the femur has been prepared (see cor respond<br />

ing surgical procedure).<br />

Outer section<br />

Disassemble the inserter (130818).<br />

Screw the corresponding size of trial insert<br />

(standard and hooded) into the inner section<br />

of the seating instrument.<br />

Insert the trial insert into the implanted shell.<br />

Unscrew the inner part of the inserter. Check<br />

the seating of the insert in the shell, reduce<br />

and manipulate.<br />

Check the correct positioning of the PE insert<br />

within the shell.<br />

Screw the internal part of the seating instrument<br />

back into the trial insert, then loosen<br />

and remove it through a gentle tilting action.<br />

Thoroughly rinse the implanted shell and carefully remove the bone debris.<br />