Optionally: trial reduction using trial<br />
insert (standard 130826–130839/hooded<br />
130850–130857)<br />
Inner section<br />
The trial stem is inserted in the femur after<br />
the femur has been prepared (see cor respond<br />
ing surgical procedure).<br />
Outer section<br />
Disassemble the inserter (130818).<br />
Screw the corresponding size of trial insert<br />
(standard and hooded) into the inner section<br />
of the seating instrument.<br />
Insert the trial insert into the implanted shell.<br />
Unscrew the inner part of the inserter. Check<br />
the seating of the insert in the shell, reduce<br />
and manipulate.<br />
Check the correct positioning of the PE insert<br />
within the shell.<br />
Screw the internal part of the seating instrument<br />
back into the trial insert, then loosen<br />
and remove it through a gentle tilting action.<br />
Thoroughly rinse the implanted shell and carefully remove the bone debris.<br />