Multi Gas Leak Detector <strong>MGLD.3000</strong><br />

Operating instructions<br />

The following error messages can be output:<br />

Error code Error message Cause / recommendations<br />

6101 No Probe present No probe was connected to the base unit.<br />

Switch off the device. Check the connection on the electropneumatic<br />

connection line on the base unit and on the probe.<br />

Check the connection line for damage such as kinks or cracks.<br />

Switch the device back on.<br />

6201 Probe: Heating Sensor The hand probe is initialized and the sensor is heated. This warning<br />

message will disappear when the sensor has finished heating<br />

(approx. 1-2 min.). If the warning message is still displayed after<br />

2 min., restart the device.<br />

6202 Probe: Reading Calibration<br />

Data<br />

The hand probe is initialized and the calibration data is read. This<br />

message shows an error while reading the calibration data from<br />

the probe. To correct the error, restart the device.<br />

6203 Probe: Writing Calibration<br />

Data<br />

Once the test- and calibration procedure is completed, the correction<br />

factor that is calculated is saved in the probe. The message<br />

indicates that an error occurred while saving the correction<br />

factor. To ensure correct transfer of the correction factor, repeat<br />

the test- and calibration procedure.<br />

6204 Probe: Reduced Flow The current through the probe is monitored by a current sensor.<br />

6205 Probe: No Flow A current that is too low, or a complete absence of current, in the<br />

hand probe can have the following causes:<br />

• Contamination/clogging of the Teflon tip<br />

• Damage (cracks, kinks in the connection line)<br />

• Insufficient connection of the probe to the connection line<br />

After correcting the root problem, the current is re-established. It<br />

can take up to 30 sec. to determine the current.<br />

In case of error the measurement characteristics of the device<br />

can be reduced. A determination of the gas concentration is still<br />

possible, although the conversion to leak rates is incorrect.<br />