Multi Gas Leak Detector <strong>MGLD.3000</strong><br />
Operating instructions<br />
Environment and selectivity setting Menu entries Warnings / functions<br />
Select reference leak<br />
and start calibration<br />
procedure<br />
Check and calibration<br />
results<br />
Environment setting<br />
Device ready to operate<br />
Probe active<br />
unlocked<br />
Connected to power socket<br />
Battery charge status<br />
Setting the environment<br />
This dialog is used to configure the environmental function for all leak search<br />
modes.<br />
The following selection can be made:<br />
• Exterior areas (outdoors)<br />
• Hall (large air flow, e.g. from open doors)<br />
• Room (small air flow, e.g. from air-conditioners)<br />
• Lab (no air flow)<br />
The transfer functions "Outdoor", "Hall", and "Room" are only suitable<br />
for a quick leak search. For a quantitative determination of<br />
leakage level, only the "lab" transfer function is recommended.<br />
Setting the selectivity<br />
By touching and moving the slider, the selectivity in each direction is changed:<br />
• Minimum selectivity: Low selectivity in polluted environments<br />
• Maximum selectivity: high selectivity low-effected environments in<br />
The selectivity adjustment is only in the refrigerant-probe effec<br />
OK - Confirmation, apply the settings and return to the preceding existing leak<br />
detection mode.<br />
Cancel and reset to original value<br />
BAL_MGLD3000_20101001_en.doc 36-54