Multi Gas Leak Detector <strong>MGLD.3000</strong><br />

Operating instructions<br />

Environment and selectivity setting Menu entries Warnings / functions<br />

Select reference leak<br />

and start calibration<br />

procedure<br />

Check and calibration<br />

results<br />

Environment setting<br />

Device ready to operate<br />

Probe active<br />

unlocked<br />

Connected to power socket<br />

Battery charge status<br />

Setting the environment<br />

This dialog is used to configure the environmental function for all leak search<br />

modes.<br />

The following selection can be made:<br />

• Exterior areas (outdoors)<br />

• Hall (large air flow, e.g. from open doors)<br />

• Room (small air flow, e.g. from air-conditioners)<br />

• Lab (no air flow)<br />

The transfer functions "Outdoor", "Hall", and "Room" are only suitable<br />

for a quick leak search. For a quantitative determination of<br />

leakage level, only the "lab" transfer function is recommended.<br />

Setting the selectivity<br />

By touching and moving the slider, the selectivity in each direction is changed:<br />

• Minimum selectivity: Low selectivity in polluted environments<br />

• Maximum selectivity: high selectivity low-effected environments in<br />

The selectivity adjustment is only in the refrigerant-probe effec<br />

OK - Confirmation, apply the settings and return to the preceding existing leak<br />

detection mode.<br />

Cancel and reset to original value<br />

BAL_MGLD3000_20101001_en.doc 36-54