Multi Gas Leak Detector <strong>MGLD.3000</strong><br />

Operating instructions<br />

OK - Save the correction factor in the probe.<br />

Cancel and return to the previous leak search menu<br />

If the sensitivity is outside the specified tolerance limits, this is indicated by a<br />

pointer that is at either red end of the scale on the left or right. The functional<br />

test was not passed. The correction factor was set to the maximum limit.<br />

Start the new test by pressing the "Start" button<br />

Cancel and return to the previous leak search menu<br />

If the test is run several consecutive times without a passing result,<br />

the following procedure can be followed:<br />

• Make sure that the Teflon tip was being held fully in front of the<br />

gas outlet of the reference leak and that there is no air movement<br />

in the environment<br />

• Make sure that the reference leak is opened for long enough to<br />

be completely permeated by the test gas (follow manufacturer<br />

instructions)<br />

• Make sure that the pressure of the test gas is consistent with the<br />

reference leak manufacturer's specifications<br />

• Restart device and repeat test (when the device is started the<br />

sensor and the gas sensor system are initialized)<br />

• Contact technical support<br />