Multi Gas Leak Detector <strong>MGLD.3000</strong><br />
Operating instructions<br />
3/5… Number of the current test point / total number of test points<br />
and to the right of this is the maximum current measured concentration<br />
or leakage rate of the current test point (in the example above, test<br />
point no. 3 and 0.9 g/y)<br />
∑… Sum of the leakage rates or concentrations of all test points up to the<br />
last/current test point<br />
The current concentration or leakage rate of the current test point is<br />
shown in the display for the probe.<br />
• When the threshold point value is exceeded the background of the<br />
screen and the numeric indicator of the concentration or leakage rate of<br />
the test point turns red. When the cumulative threshold value is exceeded,<br />
the number indicator of the cumulative concentration or leakage<br />
rate turns red. The acoustic signal is emitted in a similar manner as in<br />
analysis mode.<br />
Probe<br />
• Start of the test sequence with the selection of the first test point and<br />
indicator for the number of the current test point and the total number of<br />
configured test points in the display (Ex.: 1 of 5).<br />
• Display indicator of the currently measured concentration (value and unit<br />
in ppm) or leakage rate (value only) of the current test point. The readiness<br />
for detection of the test point is indicated by an increasing and decreasing<br />
status button under the number indicator.<br />
• Completion of the detection of the selected test point<br />
01.05<br />
20 ppm<br />
• Selection of the next test point and indicator for the number of the current<br />
test point and the total number of configured test points in the display<br />
(Ex.: 2 of 5)<br />
• Display indicator of the currently measured concentration (value and unit<br />
in ppm) or leakage rate (value only) of the current test point<br />
• Detection of the selected test point finished<br />
02.05<br />
30 ppm<br />
• …<br />
• Selection of the last test point and indicator for the number of the current<br />
test point and the total number of configured test points in the display<br />
(Ex.: 5 of 5)<br />
• Display indicator of the currently measured concentration (value and unit<br />
in ppm) or leakage rate (value only) of the current test point<br />
• Completion of the detection of the last test point<br />
05.05<br />
30 ppm<br />
• Indicator of the results of the test sequence<br />
BAL_MGLD3000_20101001_en.doc 29-54