Multi Gas Leak Detector MGLD.3000 Operating instructions Probe • Display indicator of the currently measured concentration (value and unit ppm) or leak rate (value only). • Change to dynamic search mode Entry • Change (back) to analysis mode • As the concentration increases the tone sequence of the acoustic signal of the hand probe becomes shorter. When the threshold value is exceeded, a continuous tone is emitted. The LED trend indicator also acts as an indicator for the current concentration relative to the threshold value. Up to 22%, green LED Up to 44%, 1st yellow LED Up to 66%, 2nd yellow LED Up to 100%, 1st red LED Up to 100%, 2nd red LED Dynamic search mode Menus Warnings / functions Select leak search mode Check + calibration Device ready to operate Probe active Settings Print out log and save to external data medium Service (device status, password, contact information) unlocked Connected to power socket Battery charge status The dynamic search mode is provided for quick leak detection. The display of the leakage level is made qualitatively using an indicator with changing color (green, yellow or red with increasing leakage level). Using the sliding controller, the sensitivity can be set in 5 levels (min. - lowest sensitivity to max - highest sensitivity). The optical and acoustic signal outputs are not based on the absolute concentration or leak rate, but rather on the quantity of their change. A large change in concentration generally indicates a large leak. After detection of a large concentration increase and subsequent removal of the probe from the leakage, "AUTOZERO" can be displayed between the sliding controller and the indicator for the leakage level. The sensor returns to its zero point while this message is visible. During this time the response characteristics are limited particularly when small leakages are detected. BAL_MGLD3000_20101001_en.doc 26-54

Multi Gas Leak Detector MGLD.3000 Operating instructions The signal output in dynamic search mode is also dependent on the speed with which the sensor is approaching the leak. A very slow approach corresponds to a slow increase in concentration and accordingly a minimal change in concentration. Configuration of sensitivity • Touch screen: Touching and vertical sliding of the indicator on the sliding controller to the desired sensitivity level. • Sensor (each press of the relevant button increases or reduces the sensitivity by 1 level) When changing the sensitivity level, there is no indicator of this in the display for sensor. The configured sensitivity level can only be controlled via the touch screen. The most recent sensitivity level setting is also retained after the base unit is switched off and back on. Probe • Display indicator of the currently measured concentration (value and unit in ppm) or leak rage (value only) - the value corresponds to the value of the concentration in analysis mode. • Change to analysis mode • Change (back) to dynamic search mode • As the concentration increases the tone sequence of the acoustic signal of the hand probe becomes shorter. The LED trend indicator also acts as an indicator for the change in concentration. BAL_MGLD3000_20101001_en.doc 27-54

Multi Gas Leak Detector <strong>MGLD.3000</strong><br />

Operating instructions<br />

Probe<br />

• Display indicator of the currently measured concentration (value and unit<br />

ppm) or leak rate (value only).<br />

• Change to dynamic search mode<br />

Entry<br />

• Change (back) to analysis mode<br />

• As the concentration increases the tone sequence of the acoustic signal<br />

of the hand probe becomes shorter. When the threshold value is exceeded,<br />

a continuous tone is emitted. The LED trend indicator also acts<br />

as an indicator for the current concentration relative to the threshold<br />

value.<br />

Up to 22%, green LED<br />

Up to 44%, 1st yellow LED<br />

Up to 66%, 2nd yellow LED<br />

Up to 100%, 1st red LED<br />

Up to 100%, 2nd red LED<br />

Dynamic search mode Menus Warnings / functions<br />

Select leak search<br />

mode<br />

Check + calibration<br />

Device ready to operate<br />

Probe active<br />

Settings<br />

Print out log and<br />

save to external<br />

data medium<br />

Service (device<br />

status, password,<br />

contact information)<br />

unlocked<br />

Connected to power socket<br />

Battery charge status<br />

The dynamic search mode is provided for quick leak detection. The display of<br />

the leakage level is made qualitatively using an indicator with changing color<br />

(green, yellow or red with increasing leakage level).<br />

Using the sliding controller, the sensitivity can be set in 5 levels (min. - lowest<br />

sensitivity to max - highest sensitivity).<br />

The optical and acoustic signal outputs are not based on the absolute concentration<br />

or leak rate, but rather on the quantity of their change. A large<br />

change in concentration generally indicates a large leak.<br />

After detection of a large concentration increase and subsequent removal of<br />

the probe from the leakage, "AUTOZERO" can be displayed between the<br />

sliding controller and the indicator for the leakage level. The sensor returns to<br />

its zero point while this message is visible. During this time the response<br />

characteristics are limited particularly when small leakages are detected.<br />

BAL_MGLD3000_20101001_en.doc 26-54