CrystalSix Sensor Operating Manual<br />

Table 4-1 Troubleshooting (continued)<br />

Symptom Cause Remedy<br />

Sensor jamming<br />

Material does not adhere<br />

to the crystal.<br />

Crystal holders come<br />

loose and jam sensor<br />

during rotation. May be<br />

caused by absence of<br />

orifice in airline.<br />

Electrical contacts on<br />

carousel may be<br />

damaged if unit was<br />

counter rotated.<br />

Check the cleanliness of<br />

the crystal surface;<br />

evaporate an<br />

intermediate layer of<br />

proper material on the<br />

crystal to improve<br />

adhesion.<br />

Check for orifice, see<br />

Chapter Chapter 3.<br />

Replace contacts.<br />

IPN 074-155G<br />