CrystalSix Sensor Operating Manual<br />
Table 4-1 Troubleshooting (continued)<br />
Symptom Cause Remedy<br />
Sensor jamming<br />
Material does not adhere<br />
to the crystal.<br />
Crystal holders come<br />
loose and jam sensor<br />
during rotation. May be<br />
caused by absence of<br />
orifice in airline.<br />
Electrical contacts on<br />
carousel may be<br />
damaged if unit was<br />
counter rotated.<br />
Check the cleanliness of<br />
the crystal surface;<br />
evaporate an<br />
intermediate layer of<br />
proper material on the<br />
crystal to improve<br />
adhesion.<br />
Check for orifice, see<br />
Chapter Chapter 3.<br />
Replace contacts.<br />
IPN 074-155G<br />
4 - 12