CrystalSix Sensor Operating Manual<br />

Table 4-1 Troubleshooting (continued)<br />

Symptom Cause Remedy<br />

Crystal ceases to<br />

oscillate during<br />

deposition before it<br />

reaches its “normal life.<br />

Crystal does not oscillate<br />

or oscillates<br />

intermittently (both in<br />

vacuum and in air).<br />

Crystal oscillates in<br />

vacuum but stops<br />

oscillation after open to<br />

air.<br />

Crystal near the end of its<br />

life.<br />

Scratches or foreign<br />

particles on the crystal<br />

holder seating surface.<br />

Crystal is being hit by<br />

small droplets of molten<br />

material from the<br />

evaporation source.<br />

Damaged crystal.<br />

Deposition material<br />

built-up on edge of the<br />

crystal holder and<br />

touching the crystal.<br />

Material on crystal holder<br />

partially masking full<br />

crystal area.<br />

Defective or damaged<br />

crystal.<br />

Existence of electrical<br />

short or poor electrical<br />

contacts.<br />

Crystal was near the end<br />

of its life; opening to air<br />

causes film oxidation,<br />

which increases film<br />

stress.<br />

Excessive moisture<br />

accumulation on the<br />

crystal.<br />

Replace crystal.<br />

Replace the crystal<br />

holder (IPN<br />

750-262-G1). Clean,<br />

polish the crystal seating<br />

surface of the crystal<br />

holder.<br />

Move the sensor further<br />

away from the evaporant<br />

source.<br />

Change crystal.<br />

Clean the crystal holder.<br />

Clean crystal holder.<br />

Replace crystal.<br />

Check for electrical<br />

continuity and short in<br />

sensor cable, connector,<br />

contact springs,<br />

feedthroughs.<br />

Replace crystal<br />

Turn off cooling water to<br />

sensor before opening it<br />

to air; flow hot water<br />

through the sensor when<br />

the chamber is open.<br />

IPN 074-155G<br />

4 - 10