CrystalSix Sensor Operating Manual<br />

Figure 4-2 Alignment Pin Installation<br />

4.4.2 Disassembly Instructions<br />

In the following steps, reference Figure 1-3 on page 1-6 and Figure 4-2 on page<br />

4-6.<br />

1 Remove the two #4 fasteners that secure the actuator cover (item #15) to<br />

the top plate (item #10).<br />

2 Remove the actuator cover. Remove the heat shield (item #8) by gently<br />

pulling on the outside of the shield, the shield should snap off.<br />

3 Remove the extension spring from the pawl and actuator stem assembly<br />

(item #24).<br />

4 Remove the two #4 fasteners (item #20) that secure the bellows assembly<br />

(item #19) to the top plate (item #10).<br />

5 Carefully remove the bellows assembly. The pawl and actuator stem<br />

assembly (item #24) and the detent (item #41) will be removed with the<br />

bellows assembly.<br />

6 Using the 1/16" allen wrench, loosen the set screws (item #33) that secures<br />

the ratchet (item #22) and the stop ratchet (item #38) to the carousel<br />

assembly (item #5). Remove the ratchet (item #22) and the stop ratchet<br />

(item #38).<br />

7 Remove the three #4 fasteners (item #14) that secure the top plate (item<br />

#10) to the sensor body (item #11).<br />

8 Remove the top plate (item #10).<br />

9 Remove the Teflon washer (item #39).<br />

IPN 074-155G<br />

4 - 6