CrystalSix Sensor Operating Manual<br />
2.7 CrystalSix Sensor Requirement When Not<br />
Installed With An IC/5, XTC/2, XTC/C, IC/4,<br />
or IC/4 PLUS<br />
The CrystalSix can be manually operated with other Deposition Controllers.<br />
The user must, however, be aware of the functional requirements for operation.<br />
These include, but are not limited to:<br />
1 The solenoid assembly must be energized and de-energized twice per<br />
crystal position. Twelve pulses complete one revolution.<br />
2 The electrical pulse applied to the solenoid valve must be at least one<br />
second in duration during both the pressurization and de-pressurization<br />
phase.<br />
3 A 24 volt AC or DC supply is required to energize the solenoid assembly.<br />
Consult Figure 3-1 on page 3-3 for wiring assignments.<br />
4 The user should provide some means of determining the quantity of “good”<br />
crystals when initially loaded, and quantity of good crystals remaining at any<br />
time.<br />
5 Follow the guidelines shown in section 2.3 on page 2-7 for installation.<br />
6 You will need to keep track of the total thickness as a manually generated<br />
sum of the deposit accumulated on each crystal.<br />
IPN 074-155G<br />
2 - 17