CrystalSix Sensor Operating Manual 2.2.3 IC/4 or IC/4 PLUS Deposition Controller 1 Connect the in-vacuum sensor head cable to the feedthrough or a coax adapter (microdot/BNC). 2 Connect one end of the 6" XIU cable (IPN 755-257-G6) to the BNC connector or the feedthrough. 3 Connect the other end of the 6" XIU cable to the connector of the XIU (IPN 755-252-G1). 4 Connect one end of the 15' long XIU cable (IPN 755-258-G15) to the mating connector of the XIU. 5 Connect the other end of the XIU cable to a sensor channel at the rear of the controller. 6 Connect power to the controller and set power switch to ON. Set density at 1.00 gm/cc, and zero thickness. The display should indicate 0/ or +/-.001 KÅ. Crystal life should read from 0 to 5%. 7 Breathe heavily on the crystal. A thickness indication of 1.000 to 2.000 KÅ should appear on the display. When the moisture evaporates, the thickness indication should return to approximately zero. If the above conditions are observed, you can assume the sensor is in proper working order and may be installed. IPN 074-155G 2 - 6
CrystalSix Sensor Operating Manual 2.2.4 IC-6000 or XTC Deposition Controller NOTE: See section 2.7. 1 Connect the in-vacuum sensor head cable to the feedthrough or a coax adapter (Microdot/BNC). 2 Connect one end of the 6" OSC cable (IPN 013-070) to the receptacle on the feedthrough. 3 Connect the other end of the 6" OSC cable to the receptacle marked XTAL on the oscillator (IPN 013-001). 4 Connect the end of the oscillator source/sensor cable (IPN 013-067) to the remaining BNC receptacle on the oscillator. 5 Connect the other end of the OSC source/sensor cable to the receptacle marked OSC on the rear panel of the controller. 6 Connect power to the controller and set power switch to ON. Set density at 1.00 gm/cc, and zero the thickness. The display should indicate 0/ or +/-0.001 KÅ. Crystal life should read from 0 to 5%. 7 Breathe heavily on the crystal. A thickness indication of 1.000 to 2.000 KÅ should appear on the display. When the moisture evaporates, the thickness indication should return to approximately zero. If the above conditions are observed, you can assume the sensor is in proper working order and may be installed. 2.3 General Guidelines for CrystalSix Sensor IPN 074-155G Figure 2-3, Figure 2-4, and Figure 2-5 show the typical installation of an INFICON water-cooled crystal sensor in the vacuum process chamber. Use the illustration and the following guidelines to install your sensors for optimum performance and convenience. 2 - 7
- Page 1: O P E R A T I N G M A N U A L CRYST
- Page 4 and 5: Trademarks The trademarks of the pr
- Page 6 and 7: BUSINESS REPLY MAIL FIRST CLASS PER
- Page 9 and 10: CrystalSix Sensor Operating Manual
- Page 11 and 12: 1.1 Specifications for CrystalSix S
- Page 13 and 14: CrystalSix Sensor Operating Manual
- Page 15 and 16: CrystalSix Sensor Operating Manual
- Page 17 and 18: CrystalSix Sensor Operating Manual
- Page 19 and 20: CrystalSix Sensor Operating Manual
- Page 21: CrystalSix Sensor Operating Manual
- Page 25 and 26: CrystalSix Sensor Operating Manual
- Page 27 and 28: CrystalSix Sensor Operating Manual
- Page 29 and 30: CrystalSix Sensor Operating Manual
- Page 31 and 32: CrystalSix Sensor Operating Manual
- Page 33 and 34: CrystalSix Sensor Operating Manual
- Page 35 and 36: CrystalSix Sensor Operating Manual
- Page 37 and 38: 3.3 Electrical and Pneumatic Connec
- Page 39 and 40: CrystalSix Sensor Operating Manual
- Page 41 and 42: CrystalSix Sensor Operating Manual
- Page 43 and 44: CrystalSix Sensor Operating Manual
- Page 45 and 46: CrystalSix Sensor Operating Manual
- Page 47 and 48: CrystalSix Sensor Operating Manual
- Page 49 and 50: CrystalSix Sensor Operating Manual
- Page 51 and 52: CrystalSix Sensor Operating Manual
- Page 53 and 54: 4.6 Replacing the Bellows Assembly
- Page 55 and 56: CrystalSix Sensor Operating Manual
- Page 57 and 58: CrystalSix Sensor Operating Manual
CrystalSix Sensor Operating Manual<br />
2.2.4 IC-6000 or XTC Deposition Controller<br />
NOTE: See section 2.7.<br />
1 Connect the in-vacuum sensor head cable to the feedthrough or a coax<br />
adapter (Microdot/BNC).<br />
2 Connect one end of the 6" OSC cable (IPN 013-070) to the receptacle on<br />
the feedthrough.<br />
3 Connect the other end of the 6" OSC cable to the receptacle marked XTAL<br />
on the oscillator (IPN 013-001).<br />
4 Connect the end of the oscillator source/sensor cable (IPN 013-067) to the<br />
remaining BNC receptacle on the oscillator.<br />
5 Connect the other end of the OSC source/sensor cable to the receptacle<br />
marked OSC on the rear panel of the controller.<br />
6 Connect power to the controller and set power switch to ON. Set density at<br />
1.00 gm/cc, and zero the thickness. The display should indicate 0/ or<br />
+/-0.001 KÅ. Crystal life should read from 0 to 5%.<br />
7 Breathe heavily on the crystal. A thickness indication of 1.000 to 2.000 KÅ<br />
should appear on the display. When the moisture evaporates, the thickness<br />
indication should return to approximately zero.<br />
If the above conditions are observed, you can assume the sensor is in proper<br />
working order and may be installed.<br />
2.3 General Guidelines for CrystalSix Sensor<br />
IPN 074-155G<br />
Figure 2-3, Figure 2-4, and Figure 2-5 show the typical installation of an<br />
INFICON water-cooled crystal sensor in the vacuum process chamber. Use the<br />
illustration and the following guidelines to install your sensors for optimum<br />
performance and convenience.<br />
2 - 7