CrystalSix Sensor Operating Manual Figure 2-2 Using the Crystal Puller 2.2 Pre-installation Sensor Check Prior to installing the sensor in the vacuum system, you should make certain that it is in proper working condition by following the procedure outlined below. 2.2.1 IC/5 Deposition Controller 1 Connect the in-vacuum sensor head cable to the feedthrough or a coax adapter (microdot/BNC). 2 Connect one end of the 6" XIU cable (IPN 755-257-G6) to the BNC connector or the feedthrough. 3 Connect the other end of the 6" XIU cable to the connector of the XIU/5 (IPN 760-600-G1). 4 Connect one end of the XIU/5 cable (IPN 600-1039-Gxx) to the mating connector of the XIU/5. 5 Connect the other end of the XIU/5 cable to a sensor channel at the rear of the controller. 6 Connect power to the controller and set power switch to ON. Set density at 1.00 gm/cc, and zero thickness. The display should indicate 0/ or +/-.001 KÅ. Crystal life should read from 0 to 5%. 7 Breathe heavily on the crystal. A thickness indication of 1.000 to 2.000 KÅ should appear on the display. When the moisture evaporates, the thickness indication should return to approximately zero. If the above conditions are observed, you can assume the sensor is in proper working order and may be installed. IPN 074-155G 2 - 4
CrystalSix Sensor Operating Manual 2.2.2 XTC/2 or XTC/C Deposition Controller 1 Connect the in-vacuum sensor head cable to the feedthrough or a coax adapter (microdot/BNC). 2 Connect one end of the 6" XIU cable (IPN 755-257-G6) to the BNC connector or the feedthrough. 3 Connect the other end of the 6" XIU cable to the connector of the XIU (IPN 757-302-G1). 4 Connect one end of the XX' long XIU cable (IPN 757-303-GXX) to the mating connector of the XIU. 5 Connect the other end of the XIU cable to a sensor channel at the rear of the controller. 6 Connect power to the controller and set power switch to ON. Set density at 1.00 gm/cc, and zero thickness. The display should indicate 0/ or +/-.001 KÅ. Crystal life should read from 0 to 5%. 7 Breathe heavily on the crystal. A thickness indication of 1.000 to 2.000 KÅ should appear on the display. When the moisture evaporates, the thickness indication should return to approximately zero. If the above conditions are observed, you can assume the sensor is in proper working order and may be installed. IPN 074-155G 2 - 5
- Page 1: O P E R A T I N G M A N U A L CRYST
- Page 4 and 5: Trademarks The trademarks of the pr
- Page 6 and 7: BUSINESS REPLY MAIL FIRST CLASS PER
- Page 9 and 10: CrystalSix Sensor Operating Manual
- Page 11 and 12: 1.1 Specifications for CrystalSix S
- Page 13 and 14: CrystalSix Sensor Operating Manual
- Page 15 and 16: CrystalSix Sensor Operating Manual
- Page 17 and 18: CrystalSix Sensor Operating Manual
- Page 19: CrystalSix Sensor Operating Manual
- Page 23 and 24: CrystalSix Sensor Operating Manual
- Page 25 and 26: CrystalSix Sensor Operating Manual
- Page 27 and 28: CrystalSix Sensor Operating Manual
- Page 29 and 30: CrystalSix Sensor Operating Manual
- Page 31 and 32: CrystalSix Sensor Operating Manual
- Page 33 and 34: CrystalSix Sensor Operating Manual
- Page 35 and 36: CrystalSix Sensor Operating Manual
- Page 37 and 38: 3.3 Electrical and Pneumatic Connec
- Page 39 and 40: CrystalSix Sensor Operating Manual
- Page 41 and 42: CrystalSix Sensor Operating Manual
- Page 43 and 44: CrystalSix Sensor Operating Manual
- Page 45 and 46: CrystalSix Sensor Operating Manual
- Page 47 and 48: CrystalSix Sensor Operating Manual
- Page 49 and 50: CrystalSix Sensor Operating Manual
- Page 51 and 52: CrystalSix Sensor Operating Manual
- Page 53 and 54: 4.6 Replacing the Bellows Assembly
- Page 55 and 56: CrystalSix Sensor Operating Manual
- Page 57 and 58: CrystalSix Sensor Operating Manual
CrystalSix Sensor Operating Manual<br />
2.2.2 XTC/2 or XTC/C Deposition Controller<br />
1 Connect the in-vacuum sensor head cable to the feedthrough or a coax<br />
adapter (microdot/BNC).<br />
2 Connect one end of the 6" XIU cable (IPN 755-257-G6) to the BNC<br />
connector or the feedthrough.<br />
3 Connect the other end of the 6" XIU cable to the connector of the XIU (IPN<br />
757-302-G1).<br />
4 Connect one end of the XX' long XIU cable (IPN 757-303-GXX) to the<br />
mating connector of the XIU.<br />
5 Connect the other end of the XIU cable to a sensor channel at the rear of<br />
the controller.<br />
6 Connect power to the controller and set power switch to ON. Set density at<br />
1.00 gm/cc, and zero thickness. The display should indicate 0/ or +/-.001<br />
KÅ. Crystal life should read from 0 to 5%.<br />
7 Breathe heavily on the crystal. A thickness indication of 1.000 to 2.000 KÅ<br />
should appear on the display. When the moisture evaporates, the thickness<br />
indication should return to approximately zero.<br />
If the above conditions are observed, you can assume the sensor is in proper<br />
working order and may be installed.<br />
IPN 074-155G<br />
2 - 5