PTR 225 / PTR 225 S PTR 237 - Schoonover, Inc.
PTR 225 / PTR 225 S PTR 237 - Schoonover, Inc.
PTR 225 / PTR 225 S PTR 237 - Schoonover, Inc.
Create successful ePaper yourself
Turn your PDF publications into a flip-book with our unique Google optimized e-Paper software.
Status displays<br />
Operation (POWER)<br />
Ready to measure (ignited) READY<br />
Switching threshold (active)<br />
(<strong>PTR</strong> <strong>225</strong> S only)<br />
orange LED<br />
green LED<br />
green LED<br />
Permissible load resistance<br />
R a<br />
≥ 10 kΩ<br />
Measurement signal 0.66 V to 10 V;<br />
logarithmic, 1.333 V per decade<br />
Status signal (not ignited)<br />
0.4 V<br />
1.2.2 Measurement System<br />
Measurement system<br />
Vacuum connection<br />
detachable<br />
DN 25 KF or DN 40 CF<br />
Degassing temperature see Section 2.3.3<br />
Dead volume<br />
21 cm 3 approx.<br />
Materials in contact with the medium<br />
stainless steel, CrNi,<br />
Al 2<br />
O 3<br />
ceramics, NiFe, Ni, titanium<br />
Overpressure tolerance<br />
≤ 10 bar abs.<br />
(The limits for the flange connections must be observed)<br />
Operating voltage<br />
Ignition voltage<br />
1.2.3 Signal Output<br />
Signal output<br />
GA 09.308/4.02 - 05/97<br />
1.6 kV (current limited to < 0.5 mA)<br />
2.8 kV (current limited to < 0.5 mA)<br />
0 to 10.6 V<br />
1.2.4 Switching threshold<br />
(<strong>PTR</strong> <strong>225</strong> S only)<br />
Switching threshold<br />
Adjustment range<br />
Hysteresis<br />
Rating<br />
Error status<br />
1.2.5 Control Inputs<br />
Input resistance<br />
Relay changeover contact<br />
1·10 -9 to 1·10 -3 mbar / Torr<br />
± 30 % of the adjusted range<br />
DC 60 V, 0.5 A<br />
Resting contact when<br />
“not ignited”; “HV OFF” or “supply off”<br />
R E<br />
: 10 kΩ approx.<br />
High voltage cut-in with negative logic at Pin 7:<br />
High voltage ON<br />
at U < 2.5 V<br />
High voltage OFF<br />
at U > 4 V<br />
or<br />
5<br />
5