Repair Diagnosis / Troubleshooting Script Language Daisy The script language daisy can be used to program own test routines (scripts) in addition to the existing scripts. service_gui1 service_gui2 AGFA d-lab.3 2001-01-02 /PN 9009 3.5

Diagnosis / Troubleshooting Repair Conversions − italics for non-terminal symbols − Non-proportional fond bold for key words and characters − Comments begin with a // and are valid up to the end of the line Key Words Letter fonts: Capital and small letters are not checked (case insensitive). answer ask_int ask_string calc dump end_loop include loop msgbox parameter_file set timeout trace trace_file trace_var Grammar daisy : EOF | expr_list EOF expr_list : expression NL | expression NL expr_list expression : answer | cmd | comment | dump | include | loop_head expr_list loop_tail | msgbox var | parameter_file file | timeout int | trace | trace_file file | wait | var_ask | var_calc | var_read | var_set 3.6 2001-01-02 /PN 9009 AGFA d-lab.3

Repair<br />

Diagnosis / Troubleshooting<br />

Script Language Daisy<br />

The script language daisy can be used to program<br />

own test routines (scripts) in addition to the existing<br />

scripts.<br />

service_gui1<br />

service_gui2<br />

<strong>AGFA</strong> d-<strong>lab.3</strong> 2001-01-02 /PN 9009 3.5