AGFA d-lab.3 Digital Compact Lab

AGFA d-lab.3 Digital Compact Lab AGFA d-lab.3 Digital Compact Lab

saal.group.com
from saal.group.com More from this publisher
18.06.2014 Views

Maintenance Repair Inlet Rollers in front of the IR Dryer Clean the inlet rollers Lubricate the gears with Klüberplex IR Dryer (if required) Move the sorter aside Remove the dryer cover: turn the screws 90º with a coin Disconnect the cable from the IR dryer (1) Loosen the knurled screw (2) of the IR dryer and remove the IR dryer Loosen 2 knurled screws (3) and take the insertion out of the IR dryer Clean the rollers: the inside light rubber rollers the black inlet and outlet rollers Clean the continuous chain and lubricate with Klübersynth Clean the plastic gears and lubricate with Klüberplex dlabr064 Outlet Unit Remove the top cover of the Outlet Unit Clean the Outlet Unit Check the rollers for smooth movement and if necessary lubricate the plastic bearings and the bearing points of the mobile guide elements with Constant OY 220 Replace the plastic gears if they are worn; otherwise lubricate with Klüberplex Densitometer Area (Outlet Unit) Clean the densitometer area of the Outlet Unit Carefully clean the densitometer head (Q-Tips) Check the pressure mechanism for smooth movement 9.10 2001-01-02 /PN 9009 AGFA d-lab.3

Repair Maintenance Paper Magazine Area left and right Coupling (1): visual check, replace if necessary Check the locking lever (2): The actuator roller should not slant dlabr159 AGFA d-lab.3 2001-01-02 /PN 9009 9.11

Maintenance<br />

Repair<br />

Inlet Rollers in front of the IR Dryer<br />

Clean the inlet rollers<br />

Lubricate the gears with Klüberplex<br />

IR Dryer (if required)<br />

Move the sorter aside<br />

Remove the dryer cover: turn the screws 90º<br />

with a coin<br />

Disconnect the cable from the IR dryer (1)<br />

Loosen the knurled screw (2) of the IR dryer<br />

and remove the IR dryer<br />

Loosen 2 knurled screws (3) and take the<br />

insertion out of the IR dryer<br />

Clean the rollers:<br />

the inside light rubber rollers<br />

the black inlet and outlet rollers<br />

Clean the continuous chain and lubricate with<br />

Klübersynth<br />

Clean the plastic gears and lubricate with<br />

Klüberplex<br />

dlabr064<br />

Outlet Unit<br />

Remove the top cover of the Outlet Unit<br />

Clean the Outlet Unit<br />

Check the rollers for smooth movement and if<br />

necessary lubricate the plastic bearings and the<br />

bearing points of the mobile guide elements<br />

with Constant OY 220<br />

Replace the plastic gears if they are worn;<br />

otherwise lubricate with Klüberplex<br />

Densitometer Area (Outlet Unit)<br />

Clean the densitometer area of the Outlet Unit<br />

Carefully clean the densitometer head (Q-Tips)<br />

Check the pressure mechanism for smooth<br />

movement<br />

9.10 2001-01-02 /PN 9009 <strong>AGFA</strong> d-<strong>lab.3</strong>