Maintenance Repair Inlet Rollers in front of the IR Dryer Clean the inlet rollers Lubricate the gears with Klüberplex IR Dryer (if required) Move the sorter aside Remove the dryer cover: turn the screws 90º with a coin Disconnect the cable from the IR dryer (1) Loosen the knurled screw (2) of the IR dryer and remove the IR dryer Loosen 2 knurled screws (3) and take the insertion out of the IR dryer Clean the rollers: the inside light rubber rollers the black inlet and outlet rollers Clean the continuous chain and lubricate with Klübersynth Clean the plastic gears and lubricate with Klüberplex dlabr064 Outlet Unit Remove the top cover of the Outlet Unit Clean the Outlet Unit Check the rollers for smooth movement and if necessary lubricate the plastic bearings and the bearing points of the mobile guide elements with Constant OY 220 Replace the plastic gears if they are worn; otherwise lubricate with Klüberplex Densitometer Area (Outlet Unit) Clean the densitometer area of the Outlet Unit Carefully clean the densitometer head (Q-Tips) Check the pressure mechanism for smooth movement 9.10 2001-01-02 /PN 9009 AGFA d-lab.3
Repair Maintenance Paper Magazine Area left and right Coupling (1): visual check, replace if necessary Check the locking lever (2): The actuator roller should not slant dlabr159 AGFA d-lab.3 2001-01-02 /PN 9009 9.11
- Page 393 and 394: Repair Paper Processor Signal Lists
- Page 395 and 396: Repair Paper Processor Paper Exit P
- Page 397 and 398: Repair Paper Processor Connections
- Page 401 and 402: Repair Paper Processor Functional D
- Page 403 and 404: Repair Paper Processor Electrical S
- Page 405 and 406: Repair Paper Processor Paper Proces
- Page 407 and 408: Repair Paper Processor AC/DC Power
- Page 409 and 410: Repair Paper Processor AC Power Sup
- Page 411 and 412: Repair Paper Processor Densitometer
- Page 415 and 416: Repair Paper Processor Repair Repla
- Page 417 and 418: Repair Paper Processor Replacing Pr
- Page 419 and 420: Repair Paper Processor Temperature
- Page 421 and 422: Repair Software Contents Software S
- Page 425 and 426: Repair Software Software Structure
- Page 427 and 428: Repair Software Software Structure
- Page 429 and 430: Repair Maintenance Contents Mainten
- Page 433 and 434: Repair Maintenance Maintenance Sche
- Page 435 and 436: Repair Maintenance Maintenance Jobs
- Page 437 and 438: Repair Maintenance Negative Mask: L
- Page 439 and 440: Repair Maintenance Paper Processor
- Page 441: Repair Maintenance Cleaning the Eva
- Page 445 and 446: Repair Maintenance Consumables / Cl
- Page 447 and 448: Repair General Data and Information
- Page 451 and 452: Repair General Data and Information
- Page 453 and 454: Repair General Data and Information
- Page 455 and 456: Repair General Data and Information
- Page 457 and 458: Repair General Data and Information
- Page 459 and 460: Repair General Data and Information
- Page 461 and 462: Repair General Data and Information
- Page 463 and 464: Repair Index Index A Additional Air
- Page 465 and 466: Repair Index S (cont.) Service Func
Maintenance<br />
Repair<br />
Inlet Rollers in front of the IR Dryer<br />
Clean the inlet rollers<br />
Lubricate the gears with Klüberplex<br />
IR Dryer (if required)<br />
Move the sorter aside<br />
Remove the dryer cover: turn the screws 90º<br />
with a coin<br />
Disconnect the cable from the IR dryer (1)<br />
Loosen the knurled screw (2) of the IR dryer<br />
and remove the IR dryer<br />
Loosen 2 knurled screws (3) and take the<br />
insertion out of the IR dryer<br />
Clean the rollers:<br />
the inside light rubber rollers<br />
the black inlet and outlet rollers<br />
Clean the continuous chain and lubricate with<br />
Klübersynth<br />
Clean the plastic gears and lubricate with<br />
Klüberplex<br />
dlabr064<br />
Outlet Unit<br />
Remove the top cover of the Outlet Unit<br />
Clean the Outlet Unit<br />
Check the rollers for smooth movement and if<br />
necessary lubricate the plastic bearings and the<br />
bearing points of the mobile guide elements<br />
with Constant OY 220<br />
Replace the plastic gears if they are worn;<br />
otherwise lubricate with Klüberplex<br />
Densitometer Area (Outlet Unit)<br />
Clean the densitometer area of the Outlet Unit<br />
Carefully clean the densitometer head (Q-Tips)<br />
Check the pressure mechanism for smooth<br />
movement<br />
9.10 2001-01-02 /PN 9009 <strong>AGFA</strong> d-<strong>lab.3</strong>