AGFA d-lab.3 Digital Compact Lab
AGFA d-lab.3 Digital Compact Lab AGFA d-lab.3 Digital Compact Lab
Maintenance Repair After mounting the unit, a small amount of the same grease can be applied to the rack inside while the rack is turned by hand in order to produce an even covering film (see arrow) sp14 Wet Section If required replace the replenishment valves and the bellows of the pumps: CL+P4-3805(1/1) bellow CL+P4-1119(4/1) valve (PDM) CL+P4-6110(2/1) valve for STB (Teflon, lasts longer) If required (running noise) convert the pump: Pump type MD6 Mixer of chemistry: Lubricate all belt turnaround pulleys and belt tensioners with Constant OY 220 sp17 Automatic Water Replenishment Depending on the water hardness, lime may deposit in the tray and/or on the floater switches and impair the correct functioning of the water supply (evaporation compensation). Carefully remove lime residues from the water tanks, especially from the tank inside borders as the water creeps out there through the lime rests (capillary effect). As a result, the tank is emptied prematurely the water replenishment is increased the tank chemistry is diluted Clean the floater switches: They must move freely as otherwise the automatic water replenishment does not function correctly. sp31 9.8 2001-01-02 /PN 9009 AGFA d-lab.3
Repair Maintenance Cleaning the Evaporation Measuring Recipient Open the replenisher tank lid Check the black evaporation measuring recipient (see arrow) for lime deposits, remove the deposits, if any dlabr015 Dryer Rack Remove the dryer rack Check the foam plastic rollers for wear and replace them if necessary Thoroughly clean the large plastic roller underneath the removable rollers Check the bearing blocks and replace them if necessary Check the gears and replace them if necessary Check the chain tension and tension it if necessary Lubricate the continuous drive chain with Klübersynth Lubricate alle outer gears with Klüberplex 302B002 Transport Belt Check the transport belt and retention it, if required Print Chute (if required) Lubricate the rocker gears with Klüberplex, the bearing points of the rocker with Constant OY 220 sp09 AGFA d-lab.3 2001-01-02 /PN 9009 9.9
- Page 389 and 390: Repair Paper Processor Block Circui
- Page 393 and 394: Repair Paper Processor Signal Lists
- Page 395 and 396: Repair Paper Processor Paper Exit P
- Page 397 and 398: Repair Paper Processor Connections
- Page 401 and 402: Repair Paper Processor Functional D
- Page 403 and 404: Repair Paper Processor Electrical S
- Page 405 and 406: Repair Paper Processor Paper Proces
- Page 407 and 408: Repair Paper Processor AC/DC Power
- Page 409 and 410: Repair Paper Processor AC Power Sup
- Page 411 and 412: Repair Paper Processor Densitometer
- Page 415 and 416: Repair Paper Processor Repair Repla
- Page 417 and 418: Repair Paper Processor Replacing Pr
- Page 419 and 420: Repair Paper Processor Temperature
- Page 421 and 422: Repair Software Contents Software S
- Page 425 and 426: Repair Software Software Structure
- Page 427 and 428: Repair Software Software Structure
- Page 429 and 430: Repair Maintenance Contents Mainten
- Page 433 and 434: Repair Maintenance Maintenance Sche
- Page 435 and 436: Repair Maintenance Maintenance Jobs
- Page 437 and 438: Repair Maintenance Negative Mask: L
- Page 439: Repair Maintenance Paper Processor
- Page 443 and 444: Repair Maintenance Paper Magazine A
- Page 445 and 446: Repair Maintenance Consumables / Cl
- Page 447 and 448: Repair General Data and Information
- Page 451 and 452: Repair General Data and Information
- Page 453 and 454: Repair General Data and Information
- Page 455 and 456: Repair General Data and Information
- Page 457 and 458: Repair General Data and Information
- Page 459 and 460: Repair General Data and Information
- Page 461 and 462: Repair General Data and Information
- Page 463 and 464: Repair Index Index A Additional Air
- Page 465 and 466: Repair Index S (cont.) Service Func
Repair<br />
Maintenance<br />
Cleaning the Evaporation Measuring Recipient<br />
Open the replenisher tank lid<br />
Check the black evaporation measuring recipient<br />
(see arrow) for lime deposits, remove<br />
the deposits, if any<br />
dlabr015<br />
Dryer Rack<br />
Remove the dryer rack<br />
Check the foam plastic rollers for wear and replace<br />
them if necessary<br />
Thoroughly clean the large plastic roller underneath<br />
the removable rollers<br />
Check the bearing blocks and replace them if<br />
necessary<br />
Check the gears and replace them if necessary<br />
Check the chain tension and tension it if necessary<br />
Lubricate the continuous drive chain with<br />
Klübersynth<br />
Lubricate alle outer gears with Klüberplex 302B002<br />
Transport Belt<br />
Check the transport belt and retention it, if required<br />
Print Chute (if required)<br />
Lubricate the rocker gears with Klüberplex, the<br />
bearing points of the rocker with Constant OY<br />
220<br />
sp09<br />