Repair<br />

Printer<br />

Fibre (1)<br />

General view<br />

Front view<br />

dlabr107<br />

Top plan view<br />

dlabr157<br />

dlabr155<br />

Power meter (e.g. Ophir Nova with filter or monitor<br />

diode in the laser scanning module)<br />

2 Allan keys, size 2<br />

− Completely unscrew the small Allan screw (5)<br />

− Slacken the pressure screw on the laser side until<br />

the fibre can be pulled out<br />

− Slacken the pressure screw on the exit side until<br />

the fibre holder can be pulled out<br />

− Put the fibre in the fibre holder in such a way that<br />

the pin on the fastening ring (4, no fig.) fits in the<br />

corresponding recess in the fibre holder<br />

− Mount the fibre holder with the fibre in the new<br />

laser<br />

− Tighten the pressure screws:<br />

1. Up to the stop position<br />

2. Slacken one turn<br />

dlabr156<br />

<strong>AGFA</strong> d-<strong>lab.3</strong> 2001-01-02 /PN 9009 6.4.49