Repair Printer Adjustment of the video voltages: − Adjustment of the video voltages must be made prior to the determination of AOM characteristic curves. − Execute script Scripts/scripts for service/ printer/laser/register_mode_on.txt (Answer the prompt – Switch OFF dynamic noise reduction of laser? – with YES) − Accept the suggested values (0,0,0,0,100,5000) − Connect the Multimeter at the respective point and the plug housing (GND) (Illustration lower circles) − Adjust the video voltages with the respective controls (Illustration upper circles) − Adjust the following values: red 390mV +/- 30mV green, blue 365mV +/- 30mV − Execute the script Scripts/scripts for service/ printer/laser/register_mode_on.txt. (Answer the prompt – Switch OFF dynamic noise reduction of laser? – with YES) Final check and tests Checking the software versions: − Check the firmware version in the Service GUI, compare as explained in the procedure above. Check the setup values: − Start the script Scripts/scripts for service/setup/ printer_setup_backup.txt. AGFA d-lab.3 2001-01-02 /PN 9009 6.4.9

Printer Repair − Print a test print. To do so execute Scripts/Scripts for service/ testprints/X-print.txt. Evaluate the parameter (BOL, pixel shift, center of laser line...) and correct if necessary. (In case of correction save the SETUP values again.) − Verify that jumper J5 of the Printer Mainboard is open (risk of light reflection). − For safety execute script scripts for service/initialisierung/ pr_board_selftest.txt again. − Run a communication test between the Printer Mainboard and the VIPS Board by executing the script: Script Files/scripts_Printer/ printer_general/test lvds short.txt Message at the end: Test ready /0 Error(s), LVDS-Interface o.K. − After end of this script the machine program must be started again with the ON button. − Statistics values laser: After replacement of the Printer Mainboard the runtime statistics of the laser must be started again. To do so execute the script: Scripts for service\printer\laser statistic_laser.txt Answer the questions with YES so the values for the gas laser are reset to zero. Counting of the number of ignitions and operating hours starts again. 6.4.10 2001-01-02 /PN 9009 AGFA d-lab.3

Repair<br />

Printer<br />

Adjustment of the video voltages:<br />

− Adjustment of the video voltages must be made<br />

prior to the determination of AOM characteristic<br />

curves.<br />

− Execute script Scripts/scripts for<br />

service/<br />

printer/laser/register_mode_on.txt<br />

(Answer the prompt – Switch OFF dynamic noise<br />

reduction of laser? – with YES)<br />

− Accept the suggested values (0,0,0,0,100,5000)<br />

− Connect the Multimeter at the respective point<br />

and the plug housing (GND) (Illustration lower<br />

circles)<br />

− Adjust the video voltages with the respective<br />

controls (Illustration upper circles)<br />

− Adjust the following values:<br />

red 390mV +/- 30mV<br />

green, blue 365mV +/- 30mV<br />

− Execute the script Scripts/scripts for<br />

service/<br />

printer/laser/register_mode_on.txt.<br />

(Answer the prompt – Switch OFF dynamic noise<br />

reduction of laser? – with YES)<br />

Final check and tests<br />

Checking the software versions:<br />

− Check the firmware version in the Service GUI,<br />

compare as explained in the procedure above.<br />

Check the setup values:<br />

− Start the script Scripts/scripts for<br />

service/setup/<br />

printer_setup_backup.txt.<br />

<strong>AGFA</strong> d-<strong>lab.3</strong> 2001-01-02 /PN 9009 6.4.9