Repair<br />

Scanner<br />

Scanner Calibration<br />

The following calibration procedures are performed<br />

automatically when required. It is not necessary<br />

that a technician or the operator intervenes.<br />

The Scanner contains several optical elements<br />

which are calibrated automatically.<br />

− Replacement of the light sensors in the lamphouse:<br />

The film drive software automatically finds the<br />

exact positions of the 3 light sensors in the<br />

lamphouse and the light sensor of the KG segments<br />

when the corresponding routines are<br />

called up. The positions are stored internally in<br />

an EEPROM.<br />

− White balance:<br />

Red, green, blue scan without a film to check<br />

the balance.<br />

Brightness changes, above all of the lamp<br />

brightness, are compensated.<br />

− CCD pixel correction<br />

Calibration Procedures, Correlations<br />

Performed upon:<br />

Calibration<br />

Pixel<br />

correction<br />

White<br />

balance<br />

IR adjustment<br />

Shutter<br />

damping<br />

Prescan<br />

Switch-on X X X X X<br />

Lamp change X X X X<br />

Mirror box inserted X X X X<br />

Negative carrier inserted (X) X X X<br />

Prescanner calibrated X X<br />

CCD dark reference defined X X X<br />

− X:<br />

The calibrations marked by an X must be performed<br />

when the corresponding component is<br />

modified.<br />

− (X):<br />

Required when a negative carrier is inserted for<br />

a different frame size.<br />

<strong>AGFA</strong> d-<strong>lab.3</strong> 2001-01-02 /PN 9009 4.4.27