Scanner Repair − 2 mm Allen key long (approx. 23 cm) − Open the front door − Adjust the scanner: The screw can be reached through the hole (1) with the long Allen key If the Prescanner can only be set in tolerance when the scanner tray is at the back stop, the cone shaft unit (single frame positioning) must be replaced. As this unit is adjusted as a complete unit, it can only be replaced completely (part no. on demand). dlabr171 dlabr248 Adjustment Film, Scope of Delivery Pcs. Designation Part no. 1 Test film with 2 DX tracks CM+9.8506.9931.1 1 Allen key CM+9.9999.1036.0 1 Instruction The attached Allen key cannot be used for the adjustment of the d-lab.3. 4.4.26 2001-01-02 /PN 9009 AGFA d-lab.3

Repair Scanner Scanner Calibration The following calibration procedures are performed automatically when required. It is not necessary that a technician or the operator intervenes. The Scanner contains several optical elements which are calibrated automatically. − Replacement of the light sensors in the lamphouse: The film drive software automatically finds the exact positions of the 3 light sensors in the lamphouse and the light sensor of the KG segments when the corresponding routines are called up. The positions are stored internally in an EEPROM. − White balance: Red, green, blue scan without a film to check the balance. Brightness changes, above all of the lamp brightness, are compensated. − CCD pixel correction Calibration Procedures, Correlations Performed upon: Calibration Pixel correction White balance IR adjustment Shutter damping Prescan Switch-on X X X X X Lamp change X X X X Mirror box inserted X X X X Negative carrier inserted (X) X X X Prescanner calibrated X X CCD dark reference defined X X X − X: The calibrations marked by an X must be performed when the corresponding component is modified. − (X): Required when a negative carrier is inserted for a different frame size. AGFA d-lab.3 2001-01-02 /PN 9009 4.4.27

Scanner<br />

Repair<br />

− 2 mm Allen key long (approx. 23 cm)<br />

− Open the front door<br />

− Adjust the scanner:<br />

The screw can be reached through the hole (1)<br />

with the long Allen key<br />

If the Prescanner can only be set in tolerance<br />

when the scanner tray is at the back stop, the<br />

cone shaft unit (single frame positioning) must be<br />

replaced.<br />

As this unit is adjusted as a complete unit, it can<br />

only be replaced completely (part no. on demand).<br />

dlabr171<br />

dlabr248<br />

Adjustment Film, Scope of Delivery<br />

Pcs. Designation Part no.<br />

1 Test film with 2 DX tracks CM+9.8506.9931.1<br />

1 Allen key CM+9.9999.1036.0<br />

1 Instruction<br />

The attached Allen key cannot be used for the<br />

adjustment of the d-<strong>lab.3</strong>.<br />

4.4.26 2001-01-02 /PN 9009 <strong>AGFA</strong> d-<strong>lab.3</strong>