Scanner Repair Negative Carrier Removing the Negative Carrier Some parts, e.g. the film drive and rocker motors, are only accessible after removal of the negative carrier. Screwdriver Caution! Do not unscrew the red lacquer-sealed banjo bolts. − Unscrew the marked screws (arrows) − Lift off the negative carrier #negcar01 1 Banjo bolt: Do not touch! 2 Active film exit 4.4.14 2001-01-02 /PN 9009 AGFA d-lab.3

Repair Scanner Changing the Lifting Magnet, toothed Belt and auxiliary Drive Shaft Allen key Screwdriver Lifting magnet: − Unscrew the marked screws (grub screws) − Pull off the magnet Belt: − Pull off the additional gear incl. the armature plate − Change the belt Auxiliary drive shaft: − Unscrew the holder (3 screws) − Pull out the shaft; the bearings can be removed after the lock washer has been removed − Mind the correct position of the holder when mounting it again (see Figure) dlabr247 1 Magnet 2 Armature plate 3 Gear 4 Holder 5 Auxiliary drive shaft 6 Lock washer AGFA d-lab.3 2001-01-02 /PN 9009 4.4.15

Scanner<br />

Repair<br />

Negative Carrier<br />

Removing the Negative Carrier<br />

Some parts, e.g. the film drive and rocker motors,<br />

are only accessible after removal of the negative<br />

carrier.<br />

Screwdriver<br />

Caution!<br />

Do not unscrew the red lacquer-sealed banjo<br />

bolts.<br />

− Unscrew the marked screws (arrows)<br />

− Lift off the negative carrier<br />

#negcar01<br />

1 Banjo bolt: Do not touch!<br />

2 Active film exit<br />

4.4.14 2001-01-02 /PN 9009 <strong>AGFA</strong> d-<strong>lab.3</strong>