Repair<br />

Scanner<br />

Shutter<br />

Toothed belt:<br />

Visual check: Must not brush<br />

Toothed belt pulley on the motor:<br />

Starting torque of the clamp<br />

1,8 Nm ± 0,1 Nm<br />

Toothed belt pulleys (visual check)<br />

In alignment<br />

Height mismatch ≤ 1 mm<br />

Toothed belt:<br />

Belt tension 191 Hz ± 10%<br />

Alternative F = 1,1 ± 0,1 N<br />

with 2 mm deflection<br />

Belt tension meter or meter<br />

Calliper gauge<br />

− Check the belt tension with a belt tension meter<br />

− Check the belt deflection<br />

dlabr222<br />

1 Toothed belt<br />

2 Steel core cut<br />

3 Markings (must point towards each other)<br />

4 Measuring head<br />

dlabr223<br />

1 Toothed belt<br />

2 Steel core cut<br />

3 Markings (must point towards each other)<br />

<strong>AGFA</strong> d-<strong>lab.3</strong> 2001-01-02 /PN 9009 4.4.3