68 Revised 7/22/09 (Gentry/Ranallo)<br />

Scan Factors: (For Part 2 Only - With Contrast)<br />

Smart Prep<br />

Prep Over<br />

mA<br />

Monitoring Delay<br />

(sec)<br />

Monitoring<br />

ISD (sec)<br />

Enhancement<br />

Threshold<br />

Diagnostic Delay<br />

(sec)<br />

Aortic Arch 40 10.0 2.0 50 15.0<br />

Temporal Bone: (W/O & W/ Contrast)<br />

Pediatric Child (3 – 6 yr)<br />

Scanner GE LS Xtra GE LS 16 GE LS 16 Pro GE LS V<strong>CT</strong> 64 GE LS 8<br />

Scan Type Helical Helical Helical Helical Helical<br />

Rotation Time (sec) 0.7 0.9 0.6 0.6<br />

Detector Coverage (mm)<br />

Beam Collimation (mm)<br />

10 10 10 20<br />

Detector Rows 16 16 16 16<br />

Pitch 0.562 0.562 0.562 0.531<br />

Speed (mm/rot) 5.625 5.625 5.625 10.62<br />

Detector Configuration 16 x 0.625 16 x 0.625 16 x 0.625 64 x 0.625<br />

Slice Thickness (mm) 1.25 1.25 1.25 1.25<br />

Interval (mm) 0.75 0.75 0.75 0.625<br />

Scan FOV Head Head Head Small Head<br />

kV 80 80 80 80<br />

Smart mA/ Auto mA Range 180-675 110-400 170-675 160-640<br />

Noise Index 7.9 7.9 7.9 7.9<br />

(Manual mA) 550 340 510 480<br />

Recon 1:<br />

DFOV 20 20 20 20<br />

Recon Type Standard Standard Standard Standard<br />

WW/ WL 400/40 400/40 400/40 400/40<br />

Recon Option Plus Plus Plus Full<br />

Recon Option<br />

Contrast (Part 2 Only)<br />

<strong>CT</strong> 1 <strong>CT</strong> 2 <strong>CT</strong> 3 <strong>CT</strong> 4 & ER <strong>CT</strong> East & RP <strong>CT</strong><br />

IQ Enhance