30 Revised 7/22/09 (Gentry/Ranallo) Orbit: Routine Pediatric Child (3 – 6 yr) Scanner GE LS Xtra GE LS 16 GE LS 16 Pro GE LS VCT 64 GE LS 8 Scan Type Helical Helical Helical Helical Helical Rotation Time (sec) 0.6 0.7 0.5 0.5 0.8 Detector Coverage (mm) Beam Collimation (mm) 10 10 10 20 10 Detector Rows 16 16 16 16 8 Pitch 0.562 0.562 0.562 0.531 0.625 Speed (mm/rot) 5.625 5.625 5.625 10.62 6.25 Detector Configuration 16 x 0.625 16 x 0.625 16 x 0.625 64 x 0.625 8 x 1.25 Slice Thickness (mm) 1.25 1.25 1.25 1.25 1.25 Interval (mm) 0.75 0.75 0.75 0.625 0.75 Scan FOV Head Head Head Small Head Head kV 100 100 100 100 100 Smart mA/ Auto mA Range 140-560 100-390 130-540 120-510 100-380 Noise Index 5.2 5.2 5.2 5.2 5.2 (Manual mA) 420 290 400 380 290 Recon 1: DFOV 18 18 18 18 18 Recon Type Standard Standard Standard Standard Standard WW/ WL 300/0 300/0 300/0 300/0 300/0 Recon Option Plus Plus Plus Full Plus Recon Option Recon 2: IQ Enhance DFOV 18 18 18 18 18 Recon Type Bone Plus Bone Plus Bone Plus Bone Plus Bone Plus WW/ WL 3000/300 3000/300 3000/300 3000/300 3000/300 Recon Option Plus Plus Plus Full Plus Recon Option IQ Enhance Slice Thickness (mm) 1.25 1.25 1.25 1.25 1.25 Interval (mm) 0.75 0.75 0.75 0.625 0.75 Recon 3: DFOV 18 18 18 18 Recon Type Bone Plus Bone Plus Bone Plus Bone Plus WW/ WL 3000/300 3000/300 3000/300 3000/300 Recon Option Plus Plus Plus Full Recon Option Non-Contrast CT 1 CT 2 CT 3 CT 4 & ER CT East & RP CT IQ Enhance Slice Thickness (mm) 0.625 0.625 0.625 0.625 Interval (mm) 0.375 0.375 0.375 0.312

31 Revised 7/22/09 (Gentry/Ranallo) Smart Prep Prep Over mA Monitoring Delay (sec) Monitoring ISD (sec) Enhancement Threshold Diagnostic Delay (sec) Aortic Arch 40 10.0 2.0 50 6.0 Orbit: Routine Pediatric Child (3 – 6 yr) Scanner GE LS Xtra GE LS 16 GE LS 16 Pro GE LS VCT 64 GE LS 8 Scan Type Helical Helical Helical Helical Helical Rotation Time (sec) 0.7 0.8 0.6 0.5 0.9 Detector Coverage (mm) Beam Collimation (mm) 10 10 10 20 10 Detector Rows 16 16 16 16 8 Pitch 0.562 0.562 0.562 0.531 0.625 Speed (mm/rot) 5.625 5.625 5.625 10.62 6.25 Detector Configuration 16 x 0.625 16 x 0.625 16 x 0.625 64 x 0.625 8 x 1.25 Slice Thickness (mm) 1.25 1.25 1.25 1.25 1.25 Interval (mm) 0.75 0.75 0.75 0.625 0.75 Scan FOV Head Head Head Small Head Head kV 80 80 80 80 80 Smart mA/ Auto mA Range 160-650 110-400 150-610 170-675 110-400 Noise Index 6.0 6.0 6.0 6.0 6.0 (Manual mA) 510 340 460 510 340 Recon 1: DFOV 18 18 18 18 18 Recon Type Standard Standard Standard Standard Standard WW/ WL 300/0 300/0 300/0 300/0 300/0 Recon Option Plus Plus Plus Full Plus Recon Option Recon 2: IQ Enhance DFOV 18 18 18 18 18 Recon Type Bone Plus Bone Plus Bone Plus Bone Plus Bone Plus WW/ WL 3000/300 3000/300 3000/300 3000/300 3000/300 Recon Option Plus Plus Plus Full Plus Recon Option IQ Enhance Slice Thickness (mm) 1.25 1.25 1.25 1.25 1.25 Interval (mm) 0.75 0.75 0.75 0.625 0.75 Recon 3: DFOV 18 18 18 18 Recon Type Bone Plus Bone Plus Bone Plus Bone Plus WW/ WL 3000/300 3000/300 3000/300 3000/300 Recon Option Plus Plus Plus Full Recon Option Contrast CT 1 CT 2 CT 3 CT 4 & ER CT East & RP CT IQ Enhance Slice Thickness (mm) 0.625 0.625 0.625 0.625 Interval (mm) 0.375 0.375 0.375 0.312

31 Revised 7/22/09 (Gentry/Ranallo)<br />

Smart Prep<br />

Prep Over<br />

mA<br />

Monitoring Delay<br />

(sec)<br />

Monitoring<br />

ISD (sec)<br />

Enhancement<br />

Threshold<br />

Diagnostic Delay<br />

(sec)<br />

Aortic Arch 40 10.0 2.0 50 6.0<br />

Orbit: Routine<br />

Pediatric Child (3 – 6 yr)<br />

Scanner GE LS Xtra GE LS 16 GE LS 16 Pro GE LS V<strong>CT</strong> 64 GE LS 8<br />

Scan Type Helical Helical Helical Helical Helical<br />

Rotation Time (sec) 0.7 0.8 0.6 0.5 0.9<br />

Detector Coverage (mm)<br />

Beam Collimation (mm)<br />

10 10 10 20 10<br />

Detector Rows 16 16 16 16 8<br />

Pitch 0.562 0.562 0.562 0.531 0.625<br />

Speed (mm/rot) 5.625 5.625 5.625 10.62 6.25<br />

Detector Configuration 16 x 0.625 16 x 0.625 16 x 0.625 64 x 0.625 8 x 1.25<br />

Slice Thickness (mm) 1.25 1.25 1.25 1.25 1.25<br />

Interval (mm) 0.75 0.75 0.75 0.625 0.75<br />

Scan FOV Head Head Head Small Head Head<br />

kV 80 80 80 80 80<br />

Smart mA/ Auto mA Range 160-650 110-400 150-610 170-675 110-400<br />

Noise Index 6.0 6.0 6.0 6.0 6.0<br />

(Manual mA) 510 340 460 510 340<br />

Recon 1:<br />

DFOV 18 18 18 18 18<br />

Recon Type Standard Standard Standard Standard Standard<br />

WW/ WL 300/0 300/0 300/0 300/0 300/0<br />

Recon Option Plus Plus Plus Full Plus<br />

Recon Option<br />

Recon 2:<br />

IQ Enhance<br />

DFOV 18 18 18 18 18<br />

Recon Type Bone Plus Bone Plus Bone Plus Bone Plus Bone Plus<br />

WW/ WL 3000/300 3000/300 3000/300 3000/300 3000/300<br />

Recon Option Plus Plus Plus Full Plus<br />

Recon Option<br />

IQ Enhance<br />

Slice Thickness (mm) 1.25 1.25 1.25 1.25 1.25<br />

Interval (mm) 0.75 0.75 0.75 0.625 0.75<br />

Recon 3:<br />

DFOV 18 18 18 18<br />

Recon Type Bone Plus Bone Plus Bone Plus Bone Plus<br />

WW/ WL 3000/300 3000/300 3000/300 3000/300<br />

Recon Option Plus Plus Plus Full<br />

Recon Option<br />

Contrast<br />

<strong>CT</strong> 1 <strong>CT</strong> 2 <strong>CT</strong> 3 <strong>CT</strong> 4 & ER <strong>CT</strong> East & RP <strong>CT</strong><br />

IQ Enhance<br />

Slice Thickness (mm) 0.625 0.625 0.625 0.625<br />

Interval (mm) 0.375 0.375 0.375 0.312