CT Protocols: (Brain, ENT, Spine, Vascular) - Department of Radiology
CT Protocols: (Brain, ENT, Spine, Vascular) - Department of Radiology
CT Protocols: (Brain, ENT, Spine, Vascular) - Department of Radiology
Create successful ePaper yourself
Turn your PDF publications into a flip-book with our unique Google optimized e-Paper software.
20 Revised 7/22/09 (Gentry/Ranallo)<br />
Head: Helical Scan with Angled Axial<br />
Reformations - Pediatric Infant (0– 3 yr)<br />
Scanner GE LS Xtra GE LS 16 GE LS 16 Pro GE LS V<strong>CT</strong> 64 GE LS 8<br />
Scan Type Helical Helical Helical Helical Helical<br />
Rotation Time (sec) 0.6 0.7 0.5 0.5 0.8<br />
Detector Coverage (mm)<br />
Beam Collimation (mm)<br />
10 10 10 20 10<br />
Detector Rows 16 16 16 16 8<br />
Pitch 0.562 0.562 0.562 0.531 0.625<br />
Speed (mm/rot) 5.625 5.625 5.625 10.62 6.25<br />
Detector Configuration 16 x 0.625 16 x 0.625 16 x 0.625 64 x 0.625 8 x 1.25<br />
Slice Thickness (mm) 0.625 0.625 0.625 0.625 1.25<br />
Interval (mm) 0.375 0.375 0.375 0.312 0.75<br />
Scan FOV Head Head Head Small Head Head<br />
kV 80 80 80 80 80<br />
Smart mA/ Auto mA Range 110-570 80-390 110-550 100-520 80-390<br />
Noise Index 6.6 6.6 6.6 6.6 6.6<br />
(Manual mA) 370 260 360 340 260<br />
Recon 1:<br />
DFOV 20 20 20 20 20<br />
Recon Type Standard Standard Standard Standard Standard<br />
WW/ WL 80/25 80/25 80/25 80/25 80/25<br />
Recon Option Plus Plus Plus Full Plus<br />
Recon Option<br />
Recon 2:<br />
IQ Enhance<br />
DFOV 20 20 20 20 20<br />
Recon Type Bone Plus Bone Plus Bone Plus Bone Plus Bone Plus<br />
WW/ WL 3000/300 3000/300 3000/300 3000/300 3000/300<br />
Recon Option Plus Plus Plus Full Plus<br />
Recon Option<br />
Non-Contrast or Contrast<br />
<strong>CT</strong> 1 <strong>CT</strong> 2 <strong>CT</strong> 3 <strong>CT</strong> 4 & ER <strong>CT</strong> East & RP <strong>CT</strong><br />
IQ Enhance<br />
Slice Thickness (mm) 0.625 0.625 0.625 0.625 1.25<br />
Interval (mm) 0.375 0.375 0.375 0.312 0.75