9.4 Biological Safety Cabinet Certifications<br />
All biological safety cabinets must be certified to ensure proper operation. Certification is<br />
required:<br />
• Before a cabinet is put into service.<br />
• After a cabinet has been repaired or relocated.<br />
• After a filter has been replaced.<br />
• At least annually.<br />
This certification must be performed by a contractor that is trained to National Sanitation<br />
Foundation Standard No. 49. Please note: your EHSO does not certify biological safety<br />
cabinets.<br />
It is the responsibility of the P.I. or laboratory supervisor to ensure that biological safety<br />
cabinets within laboratories under their supervision are certified annually.<br />
9.5 Types of Biological Safety Cabinets<br />
Biological safety cabinets are divided into 3 classifications. Class I and Class II cabinets, and the<br />
total containment Class III cabinets. Class II cabinets are the most common type of cabinet used<br />
on campus.<br />
• Class I: The Class I biological safety cabinet is designed to provide personnel and<br />
environmental protection only. Unfiltered air is directed through the front opening,<br />
across the work area and out through the HEPA filter on top. This cabinet is<br />
conventionally used with a full width open front, or can be used with an attached armhole<br />
front panel with or without attached rubber gloves. Although Class I cabinets are simple<br />
and economical, and radioisotopes and some toxic chemicals can be used (if the exhaust<br />
is ducted to the outside), filtered air is not provided over the work area. These cabinets do<br />
not protect your materials from contaminants introduced from the environment or the<br />
operator.<br />
• Class II: A Class II cabinet meets the requirements for the protection of product,<br />
personnel, and the environment. The capacity to protect materials within the cabinet is<br />
provided by the flow of HEPA-filtered air over the work surface. There are four subtypes<br />
of Class II cabinets based on the construction, inflow air velocities, and exhaust systems.<br />
These cabinets can be used to manipulate low to moderate risk agents.<br />
o Class IIA1: Air, at a face velocity of 75 lfpm, is drawn into the front grille of the<br />
cabinet away from the work surface. HEPA-filtered air is directed downward<br />
over the work area. As the air approaches the work surface, the blower part of the<br />
air is directed through the front grille and the remainder through the rear grille.<br />
From a common plenum, approximately 70% of the air is recirculated to the work<br />
zone through a HEPA filter and about 30% is exhausted to the room through<br />
another HEPA filter. This cabinet is unsuitable for work that involves radioactive<br />
62