<strong>Inlet</strong> <strong>pressure</strong> (<strong>PI3C</strong> <strong>pirani</strong> <strong>gauge</strong>)<br />
D 80<br />
SYMPTOM<br />
ORIGIN / DIAGNOSTIC<br />
CONFIRMATION<br />
REMEDY<br />
Critical failure<br />
Chap. C<br />
“Error/information indicator and display”<br />
<strong>Inlet</strong> <strong>pressure</strong> display<br />
does not show<br />
atmospheric <strong>pressure</strong><br />
<strong>Inlet</strong> port not vented<br />
(<strong>Inlet</strong> vent OFF on LCD)<br />
NO<br />
Select <strong>Inlet</strong> Vent ON.<br />
<strong>Inlet</strong> <strong>PI3C</strong> <strong>gauge</strong><br />
adjustment<br />
Contact customer service.<br />
No vent<br />
(vent ON requested)<br />
Vent valve coil<br />
Check valve (2) connections<br />
and coil status<br />
Contact customer service.<br />
NO<br />
Supervisor board<br />
P0302<br />
Contact customer service.<br />
Edition 02 - February 02<br />
GB 01290 -<br />
No inlet <strong>pressure</strong><br />
display<br />
<strong>Inlet</strong> <strong>PI3C</strong> <strong>gauge</strong><br />
adjustment<br />
NO<br />
<strong>Inlet</strong> <strong>pressure</strong> <strong>gauge</strong><br />
failure (impossible to adjust it)<br />
Contact customer service.<br />
Contact customer service.<br />
Alcatel Vacuum Technology France - ASM 182 TD+ - ASM 192 T2D+ User’s Manual<br />
1/3
<strong>Inlet</strong> <strong>pressure</strong> (<strong>PI3C</strong> <strong>pirani</strong> <strong>gauge</strong>)<br />
D 80<br />
SYMPTOM<br />
ORIGIN / DIAGNOSTIC<br />
CONFIRMATION<br />
REMEDY<br />
Critical failure<br />
Chap. C<br />
“Error/information indicator and display”<br />
No inlet <strong>pressure</strong> drop<br />
at start of a cycle<br />
<strong>PI3C</strong> inlet <strong>gauge</strong> head<br />
disconnected<br />
NO<br />
Secure proper inlet <strong>PI3C</strong><br />
<strong>pirani</strong> <strong>gauge</strong> head<br />
connection.<br />
Hard vacuum test cycle<br />
unauthorized (a sound is<br />
emitted when the cycle control<br />
key is pressed)<br />
Check leak detector status:<br />
example: check that<br />
SNIFFING test is not<br />
selected.<br />
NO<br />
Fuse on the P0318 board<br />
defective<br />
Replace the fuse.<br />
NO<br />
Roughing valve defective<br />
Change the coil<br />
or the valve.<br />
Edition 02 - February 02<br />
GB 01290 -<br />
2/3 Alcatel Vacuum Technology France - ASM 182 TD+ - ASM 192 T2D+ User’s Manual
<strong>Inlet</strong> <strong>pressure</strong> (<strong>PI3C</strong> <strong>pirani</strong> <strong>gauge</strong>)<br />
D 80<br />
SYMPTOM<br />
Critical failure<br />
ORIGIN / DIAGNOSTIC<br />
CONFIRMATION<br />
Chap. C<br />
“Error/information indicator and display”<br />
REMEDY<br />
Impossible to reach<br />
High sensitivity test<br />
mode<br />
Leak detector with a<br />
blank at the inlet<br />
(a volume can be a<br />
source of outgazing).<br />
<strong>Inlet</strong> <strong>PI3C</strong> <strong>pirani</strong> <strong>gauge</strong><br />
adjustment<br />
NO<br />
Internal leak<br />
NO<br />
Dry pump ACP 20/28<br />
Contact customer service.<br />
Check the tightness of the<br />
valves, seals...<br />
Check dry pump limit<br />
<strong>pressure</strong>.<br />
NO<br />
MDP 5011/5006 HDS<br />
Check MDP 5011/5006 HDS<br />
is rotating.<br />
NO<br />
Edition 02 - February 02<br />
GB 01290 -<br />
Fuse on the P0318 board<br />
defective<br />
NO<br />
High sensitivity valve<br />
defective<br />
Replace the fuse.<br />
Change the coil<br />
or the valve.<br />
Alcatel Vacuum Technology France - ASM 182 TD+ - ASM 192 T2D+ User’s Manual<br />
3/3