Molecular pump (MDP 5011/MDP 5006 HDS) - PTB Sales
Molecular pump (MDP 5011/MDP 5006 HDS) - PTB Sales
Molecular pump (MDP 5011/MDP 5006 HDS) - PTB Sales
You also want an ePaper? Increase the reach of your titles
YUMPU automatically turns print PDFs into web optimized ePapers that Google loves.
<strong>Molecular</strong> <strong>pump</strong><br />
(<strong>MDP</strong> <strong>5011</strong>/<strong>MDP</strong> <strong>5006</strong> <strong>HDS</strong>)<br />
D 50<br />
SYMPTOM<br />
Critical failure<br />
ORIGIN / DIAGNOSTIC<br />
CONFIRMATION<br />
Chap. C<br />
“Error/information indicator and display”<br />
REMEDY<br />
<strong>Molecular</strong> <strong>pump</strong> failure<br />
<strong>Molecular</strong> <strong>pump</strong><br />
overheating<br />
NO<br />
Check the connections:<br />
unplug and plug the cables<br />
on the P0326 board and<br />
the <strong>pump</strong>.<br />
<strong>Molecular</strong> <strong>pump</strong><br />
Mechanical hard spot<br />
(does not spin freely by hand)<br />
Contact customer service.<br />
<strong>Molecular</strong> <strong>pump</strong> speed<br />
P0326<br />
Switch is in running-in<br />
NO<br />
The exhaust pressure of<br />
the molecular <strong>pump</strong> is not<br />
correct (≤ 5 mbar )<br />
Check the dry <strong>pump</strong>.<br />
Edition 02 - February 02<br />
GB 01287 -<br />
NO<br />
Leaks<br />
on lines, valves, <strong>pump</strong> or cell<br />
NO<br />
Check the tightness of the<br />
canalisations, valves, <strong>pump</strong>s<br />
and cell. (alcohol test or use<br />
of another detector).<br />
Greasing of the ball bearings<br />
molecular <strong>pump</strong> to<br />
be performed<br />
Regrease the ball bearings<br />
molecular <strong>pump</strong> according to<br />
E 40.<br />
Alcatel Vacuum Technology France - ASM 182 TD+ - ASM 192 T2D+ User’s Manual<br />
1/2
<strong>Molecular</strong> <strong>pump</strong><br />
(<strong>MDP</strong> <strong>5011</strong>/<strong>MDP</strong> <strong>5006</strong> <strong>HDS</strong>)<br />
D 50<br />
SYMPTOM<br />
ORIGIN / DIAGNOSTIC<br />
CONFIRMATION<br />
REMEDY<br />
Major fault<br />
Chap. C<br />
“Error/information indicator and display”<br />
<strong>Molecular</strong> <strong>pump</strong> speed<br />
(continued)<br />
<strong>Molecular</strong> <strong>pump</strong><br />
Mechanical hard spot (does<br />
not spin freely by hand)<br />
Contact customer service.<br />
Check the molecular<br />
<strong>pump</strong> connector<br />
<strong>Molecular</strong> <strong>pump</strong> is not<br />
connected with molecular<br />
<strong>pump</strong> controller board<br />
P0326<br />
Contact customer service.<br />
GB 01287 -<br />
Edition 02 - February 02<br />
2/2 Alcatel Vacuum Technology France - ASM 182 TD+ - ASM 192 T2D+ User’s Manual