Current in Photometer / A<br />

3,675E-07<br />

3,670E-07<br />

Estability of the Luminance Standard<br />

feeding by a stabilized DC source<br />

The light emitted<br />

by the surface of the<br />

standard is not polarised.<br />

3,665E-07<br />

3,660E-07<br />

A FOTI standard<br />

photometer was used in a<br />

photometric<br />

3,655E-07<br />

bank to<br />

calibrate the luminance of<br />

3,650E-07<br />

0 100 200 300 400 500 600 700 800 the surface of the standard<br />

Time (min)<br />

in cd m -2 , measuring the<br />

Fig. 2.- Time Stability of the Luminance Standard.<br />

illuminance on the<br />

Measured with the Photometer FOTI along 12<br />

hours<br />

photometer produced by<br />

the luminous surface at<br />

different distances. The FOTI photometer was previously calibrated against standard<br />

luminous intensity lamps.<br />

The luminance of the luminous surface of the luminance standard was measured<br />

with a final expanded uncertainly (k=2) of ± 0.7%.<br />