Here - PMOD/WRC
Here - PMOD/WRC
Here - PMOD/WRC
You also want an ePaper? Increase the reach of your titles
YUMPU automatically turns print PDFs into web optimized ePapers that Google loves.
A Simple Stray-light Correction Matrix for Array Spectrometers<br />
Y. Zong, S. W. Brown, B. C. Johnson, K. R. Lykke, and Y. Ohno<br />
National Institute of Standards and Technology, Gaithersburg, Maryland, USA.<br />
Abstract. A stray-light correction matrix has been<br />
developed for array spectrometers. By using the correction<br />
matrix, measurement errors arising from stray light are<br />
corrected by a simple, fast matrix multiplication.<br />
Validation measurements demonstrate that stray-light<br />
errors are reduced by one to two orders of magnitude, to a<br />
level less than 10 -5 of the measured signal of broad-band<br />
sources, equivalent to less than one count of the<br />
15-bit-resolution instrument. By applying the stray-light<br />
correction, the error in a spectroradiometer calibration was<br />
significantly reduced in the UV region, and the error in a<br />
LED color measurement was reduced by 0.01 in<br />
chromaticity coordinates (x, y).<br />
Introduction<br />
Array spectrometers are used in a wide range of<br />
applications in the fields of spectroradiometry,<br />
spectrophotometry, colorimetry, photometry, and optical<br />
spectroscopy due to the benefits of measurement speed,<br />
sensitivity, portability, and affordability. Array<br />
spectrometers, however, are single grating instruments, and<br />
there are intrinsic limitations to their measurement<br />
accuracy. Among several sources of errors, stray light is<br />
often the dominant source of measurement error in these<br />
instruments. Because of stray light inside the instrument,<br />
significant errors can occur when measuring light sources<br />
having dissimilar spectra from the standard source. Stray<br />
light error is serious when measuring a very low level<br />
spectral component at some wavelength while there are<br />
high level components in other wavelength regions.<br />
Currently, reliable standard sources other than tungsten<br />
lamps (and deuterium lamps for UV) are not available and<br />
errors due to stray-light are inevitable.<br />
Brown, et al., developed an iterative algorithm<br />
previously that corrects stray-light errors in a<br />
spectroradiometer's spectral responsivity calibration and a<br />
test source’s spectral distribution measurement in separate<br />
steps. In this paper, we describe a simpler and faster<br />
method using a matrix called the stray-light correction<br />
matrix. In this approach, the stray-light errors in measured<br />
signals are corrected by a simple matrix multiplication.<br />
The correction is applied to all measured raw output<br />
signals, and no distinction is made for the source being<br />
measured; i.e. whether the source is a calibration source or<br />
a test source.<br />
The stray-light correction matrix<br />
The array spectrometer is first characterized for the<br />
stray-light distribution function (SDF): the ratio of the<br />
stray-light signal to the total signal within the bandpass of<br />
a spectrometer when measuring a monochromatic spectral<br />
line source. By measuring a set of line sources covering the<br />
spectral range of the instrument, and interpolating between<br />
these line spectra, a SDF matrix is obtained. The SDF<br />
matrix is used to derive the stray-light correction matrix,<br />
and the instrument’s response for stray light is corrected by<br />
Y IB = C · Y meas , (1)<br />
where C is the stray-light correction matrix, Y meas is a<br />
column vector with the measured signals, and Y IB is a<br />
column vector with stray-light corrected signals. Note<br />
that development of matrix C is required only once, unless<br />
the imaging characteristics of the instrument change.<br />
Using Eq. 1, the stray-light correction becomes a single<br />
matrix multiplication operation, and the correction can be<br />
performed in real-time with minimal impact on acquisition<br />
speed.<br />
The results of stray-light correction<br />
The effectiveness and robustness of this stray-light<br />
correction matrix has been validated. Stray-light correction<br />
matrices have been developed for several different array<br />
spectrometers. Figure 1 shows an example validation<br />
measurement. This array spectrometer has a stray light of<br />
10 -4 for a narrow-band source measurement. The<br />
spectrometer was used to measure a green broad-band<br />
bandpass filter illuminated by a tungsten incandescent<br />
lamp. The transmittance of the filter below 420 nm is less<br />
than 10 -9 . Figure 1 shows that the stray-light contribution<br />
to the measured raw signal is significant: 10 -3 below<br />
420 nm. After applying the stray-light correction, the<br />
signal below 420 nm is reduced by 2 orders of<br />
magnitude.<br />
Relative Signal<br />
1.0E+01<br />
1.0E+00<br />
1.0E-01<br />
1.0E-02<br />
1.0E-03<br />
1.0E-04<br />
1.0E-05<br />
1.0E-06<br />
1.0E-07<br />
200 300 400 500 600 700 800<br />
Wavelength (nm)<br />
Measured Sig Corrected Sig 1 count level<br />
Figure 1. Validation result of the stray light correction for a<br />
broad-band source with a bandpass filter. Thick solid line:<br />
measured raw signals from the spectrograph; thin symbol line:<br />
stray-light corrected signals; horizontal dashed line: one-count<br />
level of the 15-bit spectrograph.<br />
The stray-light correction matrix C can also be used to<br />
correct stray-light errors in narrow-band source<br />
measurements. Figure 2 shows an example of the<br />
stray-light correction for a laser source measurement. The<br />
stray-light errors are reduced by one to two orders of<br />
magnitude in this case as well.<br />
Proceedings NEWRAD, 17-19 October 2005, Davos, Switzerland 191