Both BCA <strong>Protein</strong> <strong>Assay</strong> formulations have less protein:protein<br />

variability than the Coomassie-based assays. The color response<br />

obtained for a seven point standard curve with the standard BCA<br />

<strong>Protein</strong> <strong>Assay</strong> using BSA or BGG standards shows less than a<br />

20% variation between these two proteins (Figure 3). The<br />

Coomassie assay demonstrates >30% variation in the signal<br />

generated between BSA and BGG (Table 2, page 9). There is even<br />

less variation (