Caliper GX II System Application Evaluation - PerkinElmer
Caliper GX II System Application Evaluation - PerkinElmer Caliper GX II System Application Evaluation - PerkinElmer
Protein Titer and Purity Evaluation (MW ~ 12,000 Da) Protein (R/NR) Protein at Various Conc. TYPE INFORMATION CLASSIFICATION HERE 8
Protein Titer and Purity Evaluation (MW ~ 12,000 Da) Peak Area 9000 8000 7000 6000 5000 4000 3000 2000 1000 0 Linearity y = 4.2085x + 79.163 R 2 = 0.9928 0 500 1000 1500 2000 2500 Conc. (ug/mL) • Conc.: 50 ug/mL - 2 mg/mL • Linear over three orders of magnitude • Can function as titer assay TYPE INFORMATION CLASSIFICATION HERE 9
- Page 1 and 2: Novel Analytical Technology to Supp
- Page 3 and 4: Caliper Technologies GXII System An
- Page 5 and 6: Monoclonal Antibody Stability Sampl
- Page 7: Protein Titer and Purity Evaluation
- Page 11 and 12: Conclusion Caliper LabChip GXII Sys
Protein Titer and Purity <strong>Evaluation</strong><br />
(MW ~ 12,000 Da)<br />
Protein<br />
(R/NR)<br />
Protein<br />
at Various<br />
Conc.<br />
TYPE INFORMATION CLASSIFICATION HERE<br />
8