Operative Technique<br />
Static Locking Mode<br />
Step 1:<br />
Guided Static Locking of the<br />
Talar Screw<br />
If clinical and radiological<br />
assessment does not allow for<br />
applying tibio-talar compression, a<br />
5mm Shaft Screw should be placed<br />
in the Static position of the oblong<br />
hole. Make sure the Target Arm<br />
is locked in the “Lateral Locking”<br />
position to place the screw from<br />
the lateral side of the talus.<br />
• Insert the Tissue Protection<br />
Sleeve, Long, together with<br />
the Drill Sleeve, Long, and the<br />
Trocar, Long, into the “Talus<br />
Static” hole of the Target Arm by<br />
pressing the Safety Clip (Fig. 28).<br />
• Follow the same locking<br />
procedure as described on page<br />
14 and 15 for the insertion of the<br />
Talar Screw in dynamic position.<br />
Fig 28<br />