Operative Technique Step IV (Fig. 69) The Conical Extraction Rod is then threaded and tightend into the proximal end of the nail. The Lag Screw is extracted by anti-clockwise rotation and pulling of the Lag Screwdriver. The K-Wire must then be removed. Step V (Fig. 70a & b) An appropriate sliding hammer assembly is attached to the Extraction Rod and the nail extracted. Note: It is a useful to turn the Lag Screw Screwdriver clockwise slightly first to loosen the possibly bony ingrowth into the screw threads before turning it counter clockwise. Fig. 69 Note: As an alternative for implant extraction and even in rare case of broken implants the Stryker Implant Extraction System can be used. For details, please refer to the "Implant Extraction Set" Guide or ask Stryker Representative for further assistance. Implant Extraction Set Implant Extraction Guide Module One & Two Fig. 70a Implant Extraction Set: Quick Reference Guide (B1000055) Implant Extraction Guide (B1000057) Fig. 70b 38

Operative Technique Dealing with Special Cases In the case of a comminuted fracture, there is a tendency for the fracture to become displaced posteriorly, making it difficult to place the K-Wire into the center of the neck and head. This can be solved by lifting the nail insertion Targeting Device (Fig. 71). Alternatively, an assistant can lift up the greater trochanter manually or with a reduction spoon; or support it with a sandbag. This will maintain the neck and the femur in almost the same axis, facilitating passage of the K-Wire through the center of the neck and head. Fig. 71 The position should then be checked in both the anterior-posterior and lateral views using the image intensifier. 39

Operative Technique<br />

Step IV (Fig. 69)<br />

The Conical Extraction Rod is then<br />

threaded and tightend into the<br />

proximal end of the nail. The Lag<br />

Screw is extracted by anti-clockwise<br />

rotation and pulling of the Lag Screwdriver.<br />

The K-Wire must then be<br />

removed.<br />

Step V (Fig. 70a & b)<br />

An appropriate sliding hammer<br />

assembly is attached to the Extraction<br />

Rod and the nail extracted.<br />

Note:<br />

It is a useful to turn the Lag Screw<br />

Screwdriver clockwise slightly first<br />

to loosen the possibly bony ingrowth<br />

into the screw threads before turning<br />

it counter clockwise.<br />

Fig. 69<br />

Note:<br />

As an alternative for implant extraction<br />

and even in rare case of broken<br />

implants the <strong>Stryker</strong> Implant Extraction<br />

System can be used.<br />

For details, please refer to the<br />

"Implant Extraction Set" Guide or<br />

ask <strong>Stryker</strong> Representative for further<br />

assistance.<br />

Implant Extraction Set<br />

Implant Extraction Guide<br />

Module One & Two<br />

Fig. 70a<br />

Implant Extraction Set:<br />

Quick Reference Guide (B1000055)<br />

Implant Extraction Guide (B1000057)<br />

Fig. 70b<br />