Operative Technique<br />
End Cap Insertion<br />
It is recommended to use an End<br />
Cap to close the proximal part of the<br />
nail to prevent bone ingrowth.<br />
Leave the Screwdriver for the distal<br />
locking in place and remove the <strong>Nail</strong><br />
Holding Screw using the Ball Tip<br />
Screwdriver, Spreading Screwdriver or<br />
Strike Plate. Load the End Cap (size 0)<br />
to one of the Screwdrivers and pass<br />
the assembly through the top of the<br />
Targeting Device down into the nail.<br />
Turn the handle clockwise until it<br />
stops mechanically. Remove the<br />
Screwdriver, the distal Screwdriver<br />
and the distal sleeves and remove the<br />
Targeting Device in cranial direction.<br />
Alternatively the End Cap could also<br />
be inserted free hand after removal of<br />
the Target Device.<br />
Fig. 63<br />
End Cap assembly<br />
Fig. 64<br />
Final <strong>Nail</strong> assembly<br />