Operative Technique<br />

End Cap Insertion<br />

It is recommended to use an End<br />

Cap to close the proximal part of the<br />

nail to prevent bone ingrowth.<br />

Leave the Screwdriver for the distal<br />

locking in place and remove the <strong>Nail</strong><br />

Holding Screw using the Ball Tip<br />

Screwdriver, Spreading Screwdriver or<br />

Strike Plate. Load the End Cap (size 0)<br />

to one of the Screwdrivers and pass<br />

the assembly through the top of the<br />

Targeting Device down into the nail.<br />

Turn the handle clockwise until it<br />

stops mechanically. Remove the<br />

Screwdriver, the distal Screwdriver<br />

and the distal sleeves and remove the<br />

Targeting Device in cranial direction.<br />

Alternatively the End Cap could also<br />

be inserted free hand after removal of<br />

the Target Device.<br />

Fig. 63<br />

End Cap assembly<br />

Fig. 64<br />

Final <strong>Nail</strong> assembly<br />

35