Operative Technique<br />
To verify the correct position of the<br />
Set Screw, try to turn the Lag Screwdriver<br />
gently clockwise and counterclockwise.<br />
If it is not possible to turn<br />
the Lag Screwdriver, the Set Screw is<br />
engaged in one of the grooves. If the<br />
Lag Screw Driver still moves, recorrect<br />
the handle position and tighten the Set<br />
Screw again until it engages in one of<br />
the four grooves.<br />
After slightly tightening the Set<br />
Screw it should then be unscrewed<br />
by one quarter (¼) of a turn, until<br />
a small play can be felt at the Lag<br />
Screwdriver. This ensures a free<br />
sliding of the Lag Screw.<br />
Fig. 56<br />
Set Screw insertion<br />
Make sure that the Set Screw is still<br />
engaged in the groove by checking that<br />
it is still not possible to turn the Lag<br />
Screw with the Lag Screwdriver.<br />
Note:<br />
Do not unscrew the Set Screw more<br />
than ¼ of a turn.<br />
As an alternative, the Set Screw can be<br />
inserted using <strong>Gamma3</strong> Closed Tube<br />
Clip (see next page).<br />
If distal locking is not indicated, the<br />
End Cap should be assembled to the<br />
nail end to prevent bone ingrowth.<br />
Leaving the Lag Screwdriver in<br />
place, the <strong>Nail</strong> Holding Screw is now<br />
removed using the ball tip Screw<br />
Driver or Universal Socket Wrench<br />
and turning it counter clockwise.<br />
Insert the End Cap (size 0) using<br />
the Socket Wrench or the Ball Tip<br />
Screwdriver. The End Cap should be<br />
tightened slightly.<br />
Fig. 57<br />
Please see chapter “End Cap<br />
Insertion”.<br />
Alternatively the End Cap could also<br />
be inserted free hand after removal of<br />
the Target Device.<br />
31