Operative Technique<br />

Lag Screw Insertion<br />

The objective is to position the Lag<br />

Screw either in the center or below the<br />

center of the femoral head in the anterior-posterior<br />

view and centrally in<br />

the lateral view, to provide the<br />

best load transfer to the Lag Screw.<br />

After satisfactorily positioning the<br />

K-Wire, the required Lag Screw length<br />

is measured using the Lag Screw Ruler.<br />

Before starting to measure, ensure<br />

that the Lag Screw Guide Sleeve is still<br />

pressed firmly against the lateral cortex<br />

of the femur (Fig. 47a).<br />

Fig. 47<br />

Lag Screw length measurement<br />

Place the Lag Screw Ruler directly<br />

under the K-Wire (Fig. 48).<br />

The recommended value for the Step<br />

Drill depth and the Lag Screw length<br />

can be read directly from the Lag<br />

Screw Ruler. If the value is between<br />

markings on the scale, e. g. 97mm, it<br />

should always be rounded up to the<br />

next higher value, e. g. 100mm.<br />

Fig. 47a<br />

Warning:<br />

K-Wires are not intended for re-use.<br />

They are single use only. K-Wires<br />

may be damaged or bent during<br />

surgical procedures. If a K-Wire is<br />

re-used, it may become lodged in<br />

the drill and could be advanced into<br />

the pelvis, and may damage large<br />

blood vessels or cause other serious<br />

injuries.<br />

Fig. 48<br />

Lag Screw Length Measurement<br />