Operative Technique<br />
Lag Screw Insertion<br />
The objective is to position the Lag<br />
Screw either in the center or below the<br />
center of the femoral head in the anterior-posterior<br />
view and centrally in<br />
the lateral view, to provide the<br />
best load transfer to the Lag Screw.<br />
After satisfactorily positioning the<br />
K-Wire, the required Lag Screw length<br />
is measured using the Lag Screw Ruler.<br />
Before starting to measure, ensure<br />
that the Lag Screw Guide Sleeve is still<br />
pressed firmly against the lateral cortex<br />
of the femur (Fig. 47a).<br />
Fig. 47<br />
Lag Screw length measurement<br />
Place the Lag Screw Ruler directly<br />
under the K-Wire (Fig. 48).<br />
The recommended value for the Step<br />
Drill depth and the Lag Screw length<br />
can be read directly from the Lag<br />
Screw Ruler. If the value is between<br />
markings on the scale, e. g. 97mm, it<br />
should always be rounded up to the<br />
next higher value, e. g. 100mm.<br />
Fig. 47a<br />
Warning:<br />
K-Wires are not intended for re-use.<br />
They are single use only. K-Wires<br />
may be damaged or bent during<br />
surgical procedures. If a K-Wire is<br />
re-used, it may become lodged in<br />
the drill and could be advanced into<br />
the pelvis, and may damage large<br />
blood vessels or cause other serious<br />
injuries.<br />
Fig. 48<br />
Lag Screw Length Measurement<br />
26