Operative Technique<br />
The Lag Screw should be placed in the<br />
central position of the femoral head in<br />
the lateral view (Fig. 40).<br />
Lag Screw Positioning using<br />
the One Shot Device<br />
The One Shot Device is re commended,<br />
for establishing whether the<br />
Lag Screw is in the optimum position.<br />
This device enables correct<br />
positioning of the K-Wire for Lag<br />
Screw placement before performing<br />
lateral skin incision and opening of<br />
the lateral cortex (see Fig. 41−42a).<br />
Fig. 40<br />
nail positioned<br />
too deep<br />
Fig. 41<br />
correct position<br />
Fig. 42<br />
nail positioned<br />
too high<br />
Fig. 41a<br />
A / P view<br />
Fig. 42a<br />
Lateral view<br />
23