10.05.2014 Views

Gamma3 Trochanteric Nail 180 - Stryker

Gamma3 Trochanteric Nail 180 - Stryker

Gamma3 Trochanteric Nail 180 - Stryker

SHOW MORE
SHOW LESS

You also want an ePaper? Increase the reach of your titles

YUMPU automatically turns print PDFs into web optimized ePapers that Google loves.

Operative Technique<br />

The Lag Screw should be placed in the<br />

central position of the femoral head in<br />

the lateral view (Fig. 40).<br />

Lag Screw Positioning using<br />

the One Shot Device<br />

The One Shot Device is re commended,<br />

for establishing whether the<br />

Lag Screw is in the optimum position.<br />

This device enables correct<br />

positioning of the K-Wire for Lag<br />

Screw placement before performing<br />

lateral skin incision and opening of<br />

the lateral cortex (see Fig. 41−42a).<br />

Fig. 40<br />

nail positioned<br />

too deep<br />

Fig. 41<br />

correct position<br />

Fig. 42<br />

nail positioned<br />

too high<br />

Fig. 41a<br />

A / P view<br />

Fig. 42a<br />

Lateral view<br />

23