Operative Technique<br />
When the Gamma3 Nail has been inserted<br />
to its final depth, check the anteversion<br />
of the nail. By using of the<br />
K-Wire Clip or Closed Tube Clip with<br />
K-Wires (Fig. 39). As an alternative the<br />
Gamma3 U-Wire can be used with the<br />
K-Wire Clip or Closed Tube Clip.<br />
The K-Wire Clip or Closed Tube Clip<br />
is mounted into the slots of the Target<br />
Arm by pressing the Clip flanges together.<br />
Note:<br />
Before proceeding ensure that the<br />
Nail Holding Screw is still fully<br />
tightened.<br />
Fig. 39<br />
The Lag Screw should be placed in the<br />
central position of the femoral head in<br />
the lateral view (Fig. 40).<br />
Fig. 40<br />