Technical Details - <strong>Hoffmann</strong> ® <strong>II</strong> Compact <strong>MRI</strong><br />
<strong>MRI</strong> rods are YELLOW.<br />
All <strong>MRI</strong> clamps,<br />
couplings, and tubes<br />
have GOLD headed<br />
bolts.<br />
The <strong>Hoffmann</strong>® <strong>II</strong> Compact <strong>MRI</strong><br />
is also designed <strong>for</strong> <strong>MRI</strong> use up to 3.0<br />
Tesla. To ensure patient safety during<br />
<strong>MRI</strong> procedures and to distinguish<br />
the system from the standard Non-<br />
<strong>MRI</strong> <strong>Hoffmann</strong>® <strong>II</strong> Compact<br />
System, the <strong>Hoffmann</strong>® <strong>II</strong> Compact<br />
<strong>MRI</strong> is color-coded in ORANGE.<br />
All <strong>MRI</strong> components<br />
are marked with “<strong>MRI</strong>”.<br />
<strong>Hoffmann</strong> ® <strong>II</strong> Compact <strong>MRI</strong><br />
The <strong>MRI</strong> components have been tested according to ASTM<br />
Standards F2052, F2182, and F2213.<br />
Standard <strong>Hoffmann</strong> ® <strong>II</strong> Compact (Non <strong>MRI</strong> Safe)<br />
<strong>MRI</strong><br />
These products are ferromagnetic and/or conductive;<br />
there<strong>for</strong>e, they are not <strong>MRI</strong> Safe.<br />
The color coding scheme illustrated below must be followed<br />
to ensure correct <strong>MRI</strong> usage.<br />
4941-2-020 4941-2-200<br />
4940-2-020 4940-2-200<br />
4941-1-058 4941-1-010 4941-1-020<br />
4940-1-058 4940-1-010 4940-1-020<br />
<strong>MRI</strong> posts have a GOLD tip.<br />
4941-2-140<br />
4940-2-140<br />
5048-5-XXX<br />
5049-5-XXX<br />
<strong>MRI</strong> rods are YELLOW.<br />
These standard rods come in carbon and stainless steel. Do<br />
not mix them with the <strong>MRI</strong> System since they are not <strong>MRI</strong><br />
Safe.<br />
4941-1-100<br />
4941-0-015<br />
4940-1-100<br />
4920-0-015<br />
The <strong>Hoffmann</strong>® <strong>II</strong> Compact <strong>MRI</strong> can only be guaranteed<br />
<strong>for</strong> <strong>MRI</strong> use when using Apex® Pins to build a frame.