<strong>Operative</strong> <strong>Technique</strong><br />

End Cap Insertion<br />

After removal of the Target Device, an<br />

End Cap is used. Eight different sizes<br />

of End Caps* are available to adjust<br />

nail length and to reduce the potential<br />

for bony ingrowth into the proximal<br />

threads of the nail (Fig. 31).<br />

The End Cap is inserted with the<br />

Screwdriver Shaft and Teardrop<br />

Handle after intra-operative radiographs<br />

show satisfactory reduction<br />

and hardware implantation (Fig. 32<br />

and 33). Fully seat the End Cap to<br />

minimize the potential for loosening.<br />

Fully<br />

Threaded<br />

Standard +5mm +10mm +15mm +20mm +25mm +30mm +35mm<br />

Fig. 31<br />

Thoroughly irrigate the wound to<br />

prevent debris from remaining within<br />

the knee joint. Close the wound using<br />

standard technique.<br />

Fig. 32<br />

Fig. 33<br />

* Optional 8mm diameter End Caps are available in<br />

+5, +10 and +15mm length to facilitate insertion<br />

through the Nail Adapter (see page 27).<br />