Operative Technique Entry Point The medullary canal is opened through a superolateral plateau entry portal. The center point of the portal is located slightly medial to the lateral tibial spine as visualized on the A/P radiograph and immediately adjacent and anterior to the anterior articular margin as visualized on the true lateral radiograph. It is located lateral to the midline of the tibia by an average of 6 percent of the tibial plateau width. Radiographic confirmation of this area is essential to prevent damage to the intra-articular structure during portal placement and nail insertion (Fig. 3). L M The opening should be directed with a central orientation in relation to the medullary canal. After penetrating the cortex with the 3×285mm K‐Wire (1806-0050S), the Ø12mm Rigid Reamer (1806-2014) is used to access the medullary canal (Fig. 4). Alternatively, to penetrate the cortex, the Ø10mm Straight (1806-0045), “special order” Ø11.5mm Straight (1806- 0047), or Curved (1806-0040) Awl may be used (Fig. 5). Caution: • A more distal entry point may result in damage to the posterior cortex during nail insertion. • Guiding the Rigid Reamer over the K-Wire prior to K-Wire insertion within the Proximal Tibia will help to keep it straight while guiding the opening instrument centrally towards the canal. Do not use bent K-Wires. Note: During opening the entry portal with the Awl, dense cortex may block the tip of the Awl. An Awl Plug (1806-0032) can be inserted through the Awl to avoid penetration of bone debris into the cannulation of the Awl shaft. Fig. 4 Fig. 3 Fig. 5 10
Operative Technique Unreamed Technique If an unreamed technique is preferred, the 3×800mm Smooth Tip Guide Wire (1806-0090S) is passed through the fracture site using the Guide Wire Handle (1806-1095 and 1806- 1096) (Fig. 6). The Universal Rod (1806-0110) with Reduction Spoon (1806-0125) may be used as a fracture reduction tool to facilitate Guide Wire insertion (Fig. 7), and as a “sound” to help determine the diameter of the medullary canal. The Universal Rod is 9mm diameter. Internal rotation during insertion will aid in passing the Guide Wire down the tibial shaft. The Guide Wire should lie in the center of the metaphysis and the diaphysis in both the A/P and Lateral views to avoid offset positioning of the nail. The Guide Wire handle is removed leaving the Guide Wire in place. Fig. 6 Reamed Technique For reamed techniques, the 3×800mm Ball Tip Guide Wire (1806-0080S) is inserted through the fracture site. Except for the 8mm Tibial Nails, use of the Ball Tip Guide Wire does not require a Guide Wire exchange. The Universal Rod with Reduction Spoon may be used as a fracture reduction tool to facilitate Guide Wire insertion through the fracture site (see Fig. 7). Note: The Ball Tip at the end of the Guide Wire will stop the reamer head. Fig. 7 Reaming (Fig. 8) is commenced in 0.5mm increments until cortical contact is appreciated. Final reaming should be 1mm –1.5mm larger than the diameter of the nail to be used 11 Fig. 8
- Page 1 and 2: T2 Tibial Nailing System Operative
- Page 3 and 4: Contents Page 1. Introduction 4 Imp
- Page 5 and 6: Introduction Technical Details Nail
- Page 7 and 8: Indications, Precautions & Contrain
- Page 9: Operative Technique Patient Positio
- Page 13 and 14: Operative Technique Nail Selection
- Page 15 and 16: Operative Technique If a Guide Wire
- Page 17 and 18: Operative Technique Guided Locking
- Page 19 and 20: Operative Technique For accurate dr
- Page 21 and 22: Operative Technique Freehand Distal
- Page 23 and 24: Operative Technique Dynamic Locking
- Page 25 and 26: Operative Technique Advanced Lockin
- Page 27 and 28: Features and Benefits 8mm End Cap &
- Page 29 and 30: Ordering Information - Implants T2
- Page 31 and 32: Ordering Information - Implants T2
- Page 33 and 34: Ordering Information - Instruments
- Page 35 and 36: Ordering Information - Instruments
- Page 37 and 38: Notes 37
- Page 39 and 40: Notes 39
<strong>Operative</strong> <strong>Technique</strong><br />
Entry Point<br />
The medullary canal is opened<br />
through a superolateral plateau entry<br />
portal. The center point of the portal<br />
is located slightly medial to the lateral<br />
tibial spine as visualized on the A/P<br />
radiograph and immediately adjacent<br />
and anterior to the anterior articular<br />
margin as visualized on the true lateral<br />
radiograph. It is located lateral to<br />
the midline of the tibia by an average<br />
of 6 percent of the tibial plateau width.<br />
Radiographic confirmation of this<br />
area is essential to prevent damage to<br />
the intra-articular structure during<br />
portal placement and nail insertion<br />
(Fig. 3).<br />
L<br />
M<br />
The opening should be directed with<br />
a central orientation in relation to<br />
the medullary canal. After penetrating<br />
the cortex with the 3×285mm<br />
K‐Wire (1806-0050S), the Ø12mm<br />
Rigid Reamer (1806-2014) is used to<br />
access the medullary canal (Fig. 4).<br />
Alternatively, to penetrate the cortex,<br />
the Ø10mm Straight (1806-0045), “special<br />
order” Ø11.5mm Straight (1806-<br />
0047), or Curved (1806-0040) Awl<br />
may be used (Fig. 5).<br />
Caution:<br />
• A more distal entry point may<br />
result in damage to the posterior<br />
cortex during nail insertion.<br />
• Guiding the Rigid Reamer over<br />
the K-Wire prior to K-Wire insertion<br />
within the Proximal Tibia<br />
will help to keep it straight while<br />
guiding the opening instrument<br />
centrally towards the canal. Do<br />
not use bent K-Wires.<br />
Note:<br />
During opening the entry portal<br />
with the Awl, dense cortex may<br />
block the tip of the Awl. An Awl<br />
Plug (1806-0032) can be inserted<br />
through the Awl to avoid penetration<br />
of bone debris into the cannulation<br />
of the Awl shaft.<br />
Fig. 4<br />
Fig. 3<br />
Fig. 5<br />
10