Ordering Information – Instruments<br />

Bixcut<br />

Complete range of modular and<br />

fixed-head reamers to match<br />

surgeon preference and optimize<br />

O. R. efficiency, presented in fully<br />

sterilizable cases.<br />

Large clearance rate resulting from reduced number<br />

of reamer blades coupled with reduced length of<br />

reamer head to give effective relief of pressure and<br />

efficient removal of material.<br />

Cutting flute geometry optimized to lower pressure<br />

generation.<br />

Forward- and side-cutting face combination produces<br />

efficient material removal and rapid clearance.<br />

Double-wound shaft transmits torque effectively and<br />

with high reliability. Low-friction surface finish aids<br />

rapid debris clearance.<br />

Smaller, 6 and 8mm shaft diameters significantly<br />

reduce IM pressure.<br />

Typical Standard<br />

Reamer Ø14mm<br />

C le a r a nc e a re a :<br />

32% of cross section<br />

Bixcut<br />

Reamer Ø14mm<br />

Clearance area :<br />

59% of cross section<br />

Studies 1 have demonstrated that the<br />

pressures developed within the medullary<br />

cavity through the introduction of<br />

unreamed IMnails can be far greater<br />

than those devel oped during reaming<br />

− but this depends very much upon<br />

the design of the reamer.<br />

After a three year development study 2<br />

involving several universities, the<br />

factors that determine the pressures<br />

and temperatures developed during<br />

reaming were clearly established.<br />

These factors were applied to the de -<br />

velopment of advanced reamers that<br />

demonstrate significantly better per -<br />

form ance than the best of previous<br />

designs.<br />

Bixcut<br />

1<br />

Jan Paul M. Frolke, et al. ;<br />

Intramedullary Pressure in Reamed Femoral<br />

<strong>Nailing</strong> with Two Different Reamer Designs.,<br />

Eur. J. of Trauma, 2001 #5<br />

2<br />

Medhi Massau, et al.;<br />

Pressure Changes During Reaming with Different<br />

Parameters and Reamer Designs,<br />

Clinical Orthopaedics and Related Research<br />

Number 373, pp. 295-303, 2000<br />

30