<strong>Trillium</strong> User’s <strong>Manual</strong> <strong>35</strong><br />
3.14 Monitoring the Camera<br />
Temperature Measurement<br />
The temperature of the camera can be determined by using the<br />
verify_temperature command. This command will return the temperature<br />
of the CCD sensor in degrees Celsius. For proper operation, this value should not<br />
exceed 80°C. By default, the camera only checks the sensor temperature once at<br />
startup unless it receives the appropriate command.<br />
If the temperature exceeds 80°C, the camera will report an error. Verify that the<br />
camera is getting adequate convection cooling (i.e. the camera is not enclosed<br />
and the vent holes are not blocked etc.).<br />
Voltage Measurement<br />
The command verify_voltage checks the camera’s input voltage and internal<br />
voltages. If they are within the proper range, the camera returns Ok>. Otherwise<br />
the camera returns Fail. Note that the voltage measurement feature of the camera<br />
provides only approximate results (within 10%). They should not be used to set<br />
the applied voltage to the camera. The purpose of this test is to isolate gross<br />
problems with the supply voltages.<br />
DALSA 03-32-10119-03