3.8 Calibrating the Camera<br />

Calibration Overview<br />

This camera has the ability to calibrate itself in order to improve the color balance<br />

and image flatness. This video correction operates on a pixel-by-pixel basis and<br />

implements a two-point correction for each pixel. This correction can reduce or<br />

eliminate image distortion caused by the following factors:<br />

• Fixed Pattern Noise (FPN)<br />

• Photo Response Non Uniformity (PRNU)<br />

• Color imbalance<br />

• Lens and light source non-uniformity<br />

i<br />

For more detail<br />

on this camera’s<br />

calibration and<br />

correction<br />

algorithms, see<br />

DALSA’s<br />

“<strong>Trillium</strong> <strong>TR</strong>-31<br />

Calibration”<br />

application note,<br />

doc# 03-32-<br />

00366.<br />

This video correction can also be used to adjust the gains of the camera and to<br />

configure the color balance as required.<br />

The two point correction is implemented such that for each pixel:<br />

V output = PRNU( pixel ) * (V input - FPN( pixel ))<br />

where V output = output pixel value<br />

V input = input pixel value from the CCD<br />

PRNU( pixel) = PRNU correction coefficient for this pixel<br />

FPN( pixel ) = FPN correction coefficient for this pixel<br />

The calibration algorithm is performed in two steps. The fixed offset (FPN) is<br />

determined first by performing a calibration without any light. This calibration<br />

determines exactly how much offset to subtract per pixel in order to obtain flat<br />

output when the CCD is not exposed.<br />

The white light calibration is performed next to determine the multiplication<br />

factors required to bring each pixel to the required value (balance target) for flat,<br />

white output. The white light calibration also sets the analog gains in the camera<br />

appropriately to balance all of the channels.<br />

When performing any camera calibration, random noise must be averaged out to<br />

ensure proper sampling. Make sure the sample size for calibration is large<br />

enough to average out random noise. The factory default sample size of 64 lines<br />

is adequate for most purposes. The command to set the calibration sample size is<br />

correction_set_sample 64. The current value can be read by reading the<br />

camera parameters with get_camera_parameters.<br />

DALSA 03-32-10119-03