3.8 Calibrating the Camera<br />
Calibration Overview<br />
This camera has the ability to calibrate itself in order to improve the color balance<br />
and image flatness. This video correction operates on a pixel-by-pixel basis and<br />
implements a two-point correction for each pixel. This correction can reduce or<br />
eliminate image distortion caused by the following factors:<br />
• Fixed Pattern Noise (FPN)<br />
• Photo Response Non Uniformity (PRNU)<br />
• Color imbalance<br />
• Lens and light source non-uniformity<br />
i<br />
For more detail<br />
on this camera’s<br />
calibration and<br />
correction<br />
algorithms, see<br />
DALSA’s<br />
“<strong>Trillium</strong> <strong>TR</strong>-31<br />
Calibration”<br />
application note,<br />
doc# 03-32-<br />
00366.<br />
This video correction can also be used to adjust the gains of the camera and to<br />
configure the color balance as required.<br />
The two point correction is implemented such that for each pixel:<br />
V output = PRNU( pixel ) * (V input - FPN( pixel ))<br />
where V output = output pixel value<br />
V input = input pixel value from the CCD<br />
PRNU( pixel) = PRNU correction coefficient for this pixel<br />
FPN( pixel ) = FPN correction coefficient for this pixel<br />
The calibration algorithm is performed in two steps. The fixed offset (FPN) is<br />
determined first by performing a calibration without any light. This calibration<br />
determines exactly how much offset to subtract per pixel in order to obtain flat<br />
output when the CCD is not exposed.<br />
The white light calibration is performed next to determine the multiplication<br />
factors required to bring each pixel to the required value (balance target) for flat,<br />
white output. The white light calibration also sets the analog gains in the camera<br />
appropriately to balance all of the channels.<br />
When performing any camera calibration, random noise must be averaged out to<br />
ensure proper sampling. Make sure the sample size for calibration is large<br />
enough to average out random noise. The factory default sample size of 64 lines<br />
is adequate for most purposes. The command to set the calibration sample size is<br />
correction_set_sample 64. The current value can be read by reading the<br />
camera parameters with get_camera_parameters.<br />